Maintenance data sheet
Instruction manual ATS4, Mar-16
28
7
Maintenance data sheet
CAMAG Maintenance data sheet
ATS 4
October 2013/UB
Purpose
The maintenance data sheet informs about maintenance interval of the respective
instrument as well as the proposal for IQ/OQ interval if applicable. In addition, it
identifies consumable parts with the respective replacement cycle.
Maintenance interval
Maintenance 12 Months
IQ/OQ
12 Months
Consumable parts
Part No.
Description
Replacement cycle
Heavy/frequent/occasional use
695.0061 Plunger 10µl syringe
3/6/12 Months
695.0062 Plunger 25µl syringe
3/6/12 Months
695.0063 Plunger 100µl syringe
3/6/12 Months
695.0042 Syringe 10µl
24/24/24 Months
695.0053 Syringe 25µl
24/24/24 Months
695.0043 Syringe 100µl
24/24/24 Months
695.0046 Needle for spray on application (blue)
6/12/24 Months
695.0047 Needle for contact application (red)
6/12/24 Months
115.7450 Spray nozzle
24/24/24
720.3079 Septum puncher
6/12/24
115.7434 Tube from syringe to tower
24/24/24
695.0065 Replacement seal for syringe needle 695.0046/47,
Pkg. of 5
3 Months