Maintenance and Service
Instruction Manual, Feb-17
10
3.5
Operation advice
•
Elution of a sample–free area on the plate cleans the system from residues of the
last elution. In this manner you can also measure the plate background for later
correction
•
The positioning table must be wiped off before you put a plate or foil on it.
•
Capillaries to the detector system should be kept as short as possible
•
Increasing the flow rate results in faster elution time but larger elution volume and
a rise of pressure
•
Caution: It is not recommended to leave the ports of the valve unconnected
without protecting them from dust. Please use the included taps or similar parts to
protect the ports
Check chapter Maintenance and Service for a trouble-free handling of the instrument.
4
Maintenance and Service
4.1
Cleaning
Cleaning the instrument on a regular basis in an appropriate way assures problem-free
functionality over a long period.
A good way to perform a cleaning would be to follow the decontamination instruction
outlined below.
4.2
Decontamination
Before transportation or a longer term of not using your system, decontaminate it in an
appropriate manner. The decontamination procedure below reflects the minimal
requirements, so keep in mind that the procedure for your instrument has to be adapted
according the substances used.
Decontamination procedure:
•
Disconnect the capillaries to the LC/MS system
•
Fill a bottle with 70% Ethanol as cleaning solution and connect it to the pump. Adapt
the cleaning solution according to the used substances if necessary
•
Place a clean plate underneath the elution head
•
Perform 3 elution cycles (each 1 min). After each cycle, press the cleaning button 3 times
for 3 sec.
•
Place a lint free cloth underneath the elution head
•
Move the lever into position “Elution”
•
Run the pump for 1 min
•
Press the cleaning button 3 times for 3 seconds