Unpacking/Installation
Instruction Manual, Feb-17
4
2
Unpacking/Installation
2.1
Unpacking
Observe the environmental requirements (2.1 Installation environment) when set-ting up
the instrument.
•
From the upper compartment of the package carefully unpack all components and
accessories listed on the shipping list. Make sure the shipment is complete
•
Carefully remove the instrument from the package and place it on a table
•
Push the button
(laser cross hair on/off) down and you will see laser crosshairs
on the positioning table. If there is no cross please check the batteries
2.2
Installation environment
The place for installation must meet the following requirements:
Connections
5 bar compressed air or N2. (backside of instrument)
Short capillaries from HPLC pump to TLC-MS interface 2 and
from there to the MS
Bench space
Width 27.5 cm; Depth 45 cm; Height 30 cm
Operating
temperature
The temperature should be within a range of 15 to 35 degrees
centigrade.
Humidity
Humidity and temperature conditions must not cause
condensation.
Additional inline
filter
CAMAG strongly recommends using an additional inline filter
between the TLC-MS Interface 2 (Valve port 1) and the MS.
The porosity of the filter is depending on the usage, CAMAG
recommends to use a filter with the porosity of 0.2 µm
2.3
Installation
•
Make sure the batteries at the backside of the instrument are correctly inserted
•
Connect your LC pump and MS system with the delivered set of 10-32 (1/16”)
RheFlex® Two-Piece fittings or with compatible alternatives.
o
Port 2 from the LC pump
o
Port 1 to the MS system
•
Connect the black tube to your pressured gas supply system of 4 – 5 bar
compressed air or N2.
Your instrument is now ready to use!