TLC Scanner 4
Instruction Manual, June 2018
Page 18
6
The USER-Dialog
With the USER-Dialogs you have access to the setup parameters
Scanner Baud rate and Display Contrast of your Scanner 4. To enter
the USER-Dialog:
•
Press and hold the
RESET
key, press
. Now release
RESET
and then release
. The Scanner 4 now shows "USER-Dialog"
in the display.
New values for the parameters are saved by means of the
key;
RESET discards any change in the current parameter. The key
moves between the two functions
Baud rate
and
LCD-Contrast.
Quit the USER Dialog with the
key.
6.1
Baud rate
The Baud rate is the RS232 communication speed between com-
puter and Scanner 4. To change the Baud rate:
•
Select the function Baud rate in the USER-Dialog (by means of
the
key)
•
Use the keys
and/or
to select the desired Baud rate.
•
Press
or
to save the new value. The new Baud rate is now
valid. (recommended 19`200 baud)
The new Baud rate will be recognized automatically by the CAMAG
software at the next connection.
6.2
LCD-Contrast
The keypad display of the Scanner 4 can be adapted to the current
light conditions by means of this function in the USER-Dialog. To
change the LCD-Contrast:
•
Select the function LCD-Contrast in the USER-Dialog (by means
of the
key).
•
Use the keys
and/or
to select the desired contrast.
•
Press
or
to save the new value.
6.3
NEW HG LAMP
Use this function to automatically optimize a new Mercury lamp in
the ray path and measure the lamp emission. It may take as much
as 7 minutes to run this task.
•
In the USER-Dialog advance to the function NEW HG LAMP (by
means of the
key).
•
Use the key
to select the Hg lamp optimizing function; the
message CONFIRM:
LOAD POS
appears.
Summary of Contents for TLC SCANNER 4
Page 1: ...INSTRUCTION MANUAL TLC SCANNER 4...
Page 22: ......