TLC Scanner 4
Instruction Manual, June 2018
Page 20
Compartment
illumination
4 W fluorescent tube: 254nm (standard) or 366nm or COOL WHITE (op-
tional).
Optical system
Apochromatic Suprasil optic with 190 - 900 nm transmission. Automatic
switching between MACRO and MICRO image depending on slit size
selected.
Monochromator Grating type, band width selectable 5/20 nm, wavelength range 190 -
900 nm. Stepper motor driven, accuracy better than 1 nm.
Scanning slit
48 fixed, software selectable slit sizes ranging from 0,25x0,1 to
12x0.9mm.
Detectors
Two wide-band photomultipliers, multi-alkali type, spectral sensitivity
185 -900 nm.
Stage
movement
Independent stepper motors to drive the X and Y movement.
•
Reproducibility of position in Y = 0.05, in X = 0.1 mm.
•
Manual movement speed: 0-20 mm/s
•
Scanning speed selectable 5 - 100 mm/s, return 150 mm/s.
Power supply
115V and 230V, 50/60 Hz. Power consumption max. 180 W (Tungsten
and Mercury lamps ON)
Nitrogen
Consumption at approx. 0.5 bar: 6 L/min.
Interface
RS 232 port for connection to the PC.
Weight
37 kg
Dimensions
Width = 570 (590) mm, Depth = 590 (650) mm, height = 367 mm
7.1
UV intensity of compartment illumination
7.2
Cut-off Filter position
Wavel-
ength
2cm above TLC plate
At compartment door
µW/cm
2
MAC time eye/skin
µW/cm
2
MAC time eye/skin
254nm
650
9 sec
10
10 min
Filter
Position
320 nm
1
Empty
2
540 nm
3
400 nm
4
790 nm
5
User 1
6
User 2
7
User 3
8
CAMAG (Switzerland)
· Sonnenmattstrasse 11 · CH-4132 Muttenz 1
Telephone
+
41 61 467 34 34 · Fax
+
41 61 461 07 02 · E-Mail:
CAMAG (Germany)
· Bismarckstraße 27-29 · D-12169 Berlin
Telephone
+
49 30 516 55 50 · Fax
+
CAMAG Scientific Inc. (USA)
· 515 Cornelius Harnett Drive · Wilmington, NC 28401
Telephone 800 334 3909 · Fax 910 343 1834 · E-Mail:
www.camag.com