Avoid placing the instrument near equipment that radiates heat.
Do not place the instrument near gas burners, electric heaters or
ovens.
•
Do not place the instrument near equipment that generates in-
tense magnetic fields such as electric welding equipment, high
frequency furnaces, pole transformers, etc.
•
Protect the instrument from excessive dust.
•
Connect the instrument to power lines that are free from sudden
changes or voltage fluctuations.
•
If you must use power motor driven equipment (such as a stirrer
or shaker) in the same line as your instrument, ensure that a
noise reduction unit is in the same power line.
3.2
Conditions for the installation
Check the following requirements before installing the instrument:
Power supply and ground
Line voltage:
115 or 230VAC (make sure the voltage selector
switch is correctly set).
Frequency:
50 or 60 Hz
Power capacity: 180 W
Ground
terminal:
A grounded outlet should be within 2 m of the in-
strument.
Gas supply for flushing the monochromator (optional)
Nitrogen
Required pressure 0.5 bar (7 psi)
Gas consumption 6 L/min
3.3
Removal of the shipping protection
Open the Scanner 4 by sliding the door to the right side and remove
the knurled screws and the three brackets (1) (2) (3). Keep the
whole shipping protection set in a suitable place. Prior to shipping
the instrument the brackets must be screwed back on.