Install & Operating Manual
CleanSeal V3
Filter Validation with Mobile Integrity Probe
The Mobile Integrity Probe (MIP) is a manual scanning system for for HEPA filter validation purposes in
exhaust air applications, integrated downstream into the housing,
see also § 9.4
➔
A comprehensive testing report on system performance is available, upon request.
➔
For scanning speed and sampling rates with respect to the airflow, please refer to ISO 14644-3
or other applicable standard.
•
For reliable results and an ISO 14644-3 compliant aerosol distribution, we recommend using the
Camfil Testing Shroud,
see § 10.4.5
This accessory integrates aerosol injection and upstream concentration ports.
•
The downstream sampling port and control of the probe are located in the Control Box,
see § 9.4.1
•
The Control Box door may remain in place to deposit small parts
or it may be taken off before operation, if necessary.
IMPORTANT:
Use only the Camfil qualified MIP driver
to operate the system
(
see also § 10.4.7
DO NOT use general purpose electrical
screwdrivers !
Exceeding drive torque will damage the flexishaft and require
breaking confinement or access to the technical area for
replacement.
Mini-Latch spring (ControlBox door lock)
Pressure ports +/- for pressure drop on the filter
Manometer
Box clamping rod
Scan drive port
Sampling - Connect a particle counter or photometer
port
using G1/4 fittings RIEGLER 246.32 or 246.12
or equivalent.
Caution
Switch off the particle counter before
disconnecting as the port seals airtight
automatically on both fittings!
7