Install & Operating Manual
CleanSeal V3
Special precautions when installing a MIP housing
IMPORTANT:
Danger of damages from inappropriate unpacking or handling.
The housing is delivered fully assembled
and tested.
Do not touch the scanning system during
manipulation or installation as any deformation
or misalignment might result in functional
damage.
Protect the inside of the housing and especially the MIP support
from dirt, etc. during installation or intervention.
In case of intervention through the housing, dismantle the
probe and store in a safe location :
•
Retain probe orientation for re-installation.
•
Unplug the silicone sampling tube from the probe
•
Loosen 2 screws
to detach the probe, be careful
not to lose washers and nuts
that may fall out
of the support slots.
For re-installation, proceed as follows :
•
Insert nuts into slots
, install probe on the chariot -
align and tighten 2 screws with washers.
o
The tube connector must be on the side of the
housing sampling port fitting.
o
The probe opening must be parallel to the module
frame and at a right angle to its support rails.
•
Connect the silicone tube to its fittings.
o
The tube should always stay on the (back) side of
the scan support and must never exceed the level
of the probe opening (risk of filter damage in
application).
Rotate tube prior to connecting, so that the
natural tension keeps it on the correct side.
o
The plenum may contain a tube support bar.
If so, the tube must pass behind the support.
Please Note :
For 3P3 housings the distance is too small to avoid a
collapse of the tube with conventional hookup.
The tube needs to be installed in a loop.
No
Yes
No
Yes
7