Filter Validation for Supply Air
•
Connect a particle counter or photometer with dilution stage
to the concentration port in the housing.
•
Connect an aerosol generator to the injection port
in the housing
or
in the IntegriBox.
Caution
When using the port in the housing, switch off
the particle counter before disconnecting as the
port seals airtight automatically!
•
Upstream pressure may be monitored in parallel, using
the dp-port in the housing
and / or
in the IntegriBox.
•
Use your sampling device to scan the filter manually,
referring to applicable standards.
Filter Validation for Extract Air
For exhaust housings, please refer to the following chapters:
➔
For aerosol injection with Testing Shroud,
see § 10.4.5
➔
For filter validation with Mobile Integrity Probe,
see § 10.4.6