10.1
Input
10.1.1
Analog multifunction inputs
Number
Standard version without universal inputs.
Optional multifunction input cards (slot 1-5) each with 4 universal inputs (4/8/12/16/20).
Function
You can choose between the measured variables U, I, RTD, TC, pulse input or frequency input for
each universal input.
Measured variable, measuring
range
To IEC 60873-1:
An additional display error of ±1 digit is permitted for every measured value.
Measuring ranges which can be selected per channel:
Measured
variable
Measuring range
Maximum measured error
of measuring range (oMR)
Input
resistance
Current (I)
0 to 20 mA
0 to 5 mA
4 to 20 mA
Overrange: up to 22 mA
±0.1% oMR
Load:
50
Ohm
Voltage (U)
>1 V
0 to 10 V
0 to 5 V
1 to 5 V
±10 V
±30 V
±0.1% oMR
1 MOhm
Voltage (U)
1 V
0 to 1 V
±1 V
±150 mV
±0.1% oMR
2.5 MOhm
Resistance
thermometer
(RTD)
Pt100: -200 to 850 °C (-328 to 1562 °F) (IEC 60751, ?=0,00385)
Pt100: -200 to 850 °C (-328 to 1562 °F) (GOST 6651-94, ?=0,00391)
Pt100: -200 to 510 °C (-328 to 950 °F) (JIS C1604, ?=0,003916)
Pt500: -200 to 850 °C (-328 to 1562 °F) (IEC 60751, ?=0,00385)
Pt500: -200 to 510 °C (-328 to 950 °F) (JIS C1604, ?=0,003916)
Pt1000: -200 to 600 °C (-328 to 1112 °F) (IEC 60751, ?=0,00385)
Pt1000: -200 to 510 °C (-328 to 950 °F) (JIS C1604, ?=0,003916)
4-wire: ±0.1% oMR
3-wire: ±(0.1% oMR + 0.8 K)
2-wire: ±(0.1% oMR + 1.5 K)
Cu100: -200 to 200 °C (-328 to 392 °F) (GOST 6651-94,
α
=0,00428)
Cu50: -200 to 200 °C (-328 to 392 °F) (GOST 6651-94,
α
=0,00428)
Pt50: -200 to 850 °C (-328 to 1562 °F) (GOST 6651-94,
α
=0,00391)
4-wire: ±0.2% oMR
3-wire: ±(0.2% oMR + 0.8 K)
2-wire: ±(0.2% oMR + 1.5 K)
Cu53: -50 to 180 °C (-58 to 356 °F) (GOST 6651-94,
α
=0,00428)
Pt46: -200 to 650 °C (-328 to 1202 °F) (GOST 6651-94,
α
=0,00391)
4-wire: ±0.3% oMR
3-wire: ±(0.3% oMR + 0.8 K)
2-wire: ±(0.3% oMR + 1.5 K)