CMP6-L, CMP11-L, and CMP21-L Pyranometers
'
CR1000 Series Datalogger
Public PTemp
Public Batt_Volt
Public CMP11_Irr
Units CMP11_Irr = W/m2
DataTable (TenMin,1,-1)
DataInterval (0,1,Min,4)
Minimum (1,Batt_Volt,FP2,0,False)
Sample (1,PTemp,FP2)
Average (1,CMP11_Irr,FP2,False)
StdDev (1,CMP11_Irr,FP2,False)
EndTable
BeginProg
Scan (1,Sec,0,0)
'
Measure the Battery Voltage and Panel Temperature
PanelTemp (PTemp,250)
Battery (Batt_Volt)
'
Measure the CMP11
VoltDiff (CMP11_Irr,1,mV25C,2,True ,10000,_60Hz,1000/8.55,0)
CallTable TenMin
NextScan
EndProg
7.3.3.3 CR1000 Example Program for Measuring a CMP21
Although this example is for the CR1000, other CRBasic dataloggers are
programmed similarly. The following program measures the CMP21 every
second. It converts the pyranometer’s millivolt output to W
•
m
–2
. A
pyranometer calibration of 8.65 µV / W
•
m
–2
is used for the example program.
The resistance of the internal thermistor is converted to degree Celsius and then
to Kelvin. Every 10 minutes, the program outputs the average and standard
deviation of the flux (W
•
m
–2
) measurements and temperature measurements.
Wiring for this example is given in TABLE 7-8.
TABLE 7-8. CR1000 Wiring for CMP21 Example Program
Wire Color
Description
CR1000
Jumper*
White
Solar Signal (+)
3H
Blue
Solar Signal (–)
3L
Yellow
Voltage Excitation
VX1
Black
Temp Signal
15 SE
Brown Signal
Reference
Clear
Shield
* Jumper 3L to
with user-supplied 26 AWG or larger wire.
19
Summary of Contents for CMP11-L
Page 2: ......
Page 8: ...Table of Contents iv ...
Page 34: ...CMP6 L CMP11 L and CMP21 L Pyranometers 26 ...
Page 40: ...Appendix A CVF3 Heater Ventilator A 6 ...
Page 41: ......