Section 7. Installation
Figure 60. Strain-Gage Shunt Calibration Finish
7.9.12.6.4 FieldCalStrain() Quarter-Bridge Zero
Continuing from
FieldCalStrain() Quarter-Bridge Shunt Example
(p. 226),
keep the
249 kΩ resistor in place to simulate a strain. Using the CR1000KD Keyboard
Display or software numeric monitor, change the value in variable
Zero_Mode
to
1
to start the zero calibration as shown in figure
Zero Procedure Start
(p. 227).
When
Zero_Mode
increments to
6
, zero calibration is complete as shown in
figure
Zero Procedure Finish
(p. 227).
Figure 61. Zero Procedure Start
Figure 62. Zero Procedure Finish
227