Measurement Accuracy Example
The following example illustrates the effect percent-of-reading and offset have on
measurement accuracy. The effect of offset is usually negligible on large signals:
Example:
•
Sensor-
signal voltage: ≈2500 mV
•
CRBasic measurement instruction:
VoltDiff()
•
Programmed input-voltage range (
Range
):
mV2500
(±2500 mV)
•
Input measurement reversal (
RevDiff
):
True
•
CR1000 circuitry temperature: 10 °C
Accuracy of the measurement is calculated as follows:
accuracy = percent-of-r offset
where
percent-of-
reading = 2500 mV • ±0.06%
= ±1.5 mV
and
offset = (1.5 • 667 µV) + 1 µV
= 1.00 mV
Therefore,
accuracy = ±1.5 mV + 1.00 mV
= ±2.5 mV
Integration
The CR1000 incorporates circuitry to perform an analog integration on voltages to
be measured prior to the
A-to-D
(p. 507)
conversion. Integrating the the analog
signal removes noise that creates error in the measurement. Slow integration
removes more noise than fast integration. When the duration of the integration
matches the duration of one cycle of ac power mains noise, that noise is filtered
out. The table
Analog Measurement Integration
(p. 316)
lists valid integration
duration arguments.
Faster integration may be preferred to achieve the following objectives:
•
Minimize time skew between successive measurements
•
Maximize throughput rate
•
Maximize life of the CR1000 power supply
•
Minimize polarization of polar sensors such as those for measuring
conductivity, soil moisture, or leaf wetness. Polarization may cause
measurement errors or sensor degradation.
•
Improve accuracy of an LVDT measurement. The induced voltage in an
LVDT decays with time as current in the primary coil shifts from the inductor
to the series resistance; a long integration may result in most of signal
decaying before the measurement is complete.
Read More
See White Paper "Preventing and Attacking Measurement Noise
Problems" at
www.campbellsci.com
.
315
Summary of Contents for CR1000
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 40: ......
Page 60: ...Section 4 System Quickstart Figure 16 PC200W View Line Graph 60 ...
Page 96: ......
Page 98: ...98 ...
Page 302: ......
Page 453: ...Section 8 Operation Figure 115 Using the Keyboard Display 453 ...
Page 456: ...Section 8 Operation Figure 118 Real Time Custom 456 ...
Page 457: ...Section 8 Operation 8 8 1 3 Final Memory Tables Figure 119 Final Memory Tables 457 ...
Page 458: ...Section 8 Operation 8 8 2 Run Stop Program Figure 120 Run Stop Program 458 ...
Page 460: ...Section 8 Operation Figure 122 File Edit 460 ...
Page 461: ...Section 8 Operation 8 8 4 PCCard Memory Card Display Figure 123 PCCard CF Card Display 461 ...
Page 478: ......
Page 506: ......
Page 536: ......
Page 636: ......
Page 642: ......
Page 644: ......
Page 676: ......
Page 677: ......