Section 8. Operation
Table 75.
Calibrate()
Instruction Results
Array
Cal()
Element
Descriptions of Array Elements
Typical Value
Differential (Diff)
Single-Ended (SE)
Offset or Gain
±mV Input
Range
Integration
44
Diff
Offset
250
50 Hz Rejection
±5 LSB
45
Gain
250
50 Hz Rejection
–0.067 mV/LSB
46
SE
Offset
25
50 Hz Rejection
±5 LSB
47
Diff
Offset
25
50 Hz Rejection
±5 LSB
48
Gain
25
50 Hz Rejection
–0.0067 mV/LSB
49
SE
Offset
7.5
50 Hz Rejection
±10 LSB
50
Diff
Offset
7.5
50 Hz Rejection
±10 LSB
51
Gain
7.5
50 Hz Rejection
–0.002 mV/LSB
52
SE
Offset
2.5
50 Hz Rejection
±20 LSB
53
Diff
Offset
2.5
50 Hz Rejection
±20 LSB
54
Gain
2.5
50 Hz Rejection
–0.00067 mV/LSB
8.1.3 Pulse Measurements — Details
Related Topics:
• Pulse Measurements
— Specifications
•
Pulse Measurements — Overview
(p. 68)
•
Pulse Measurements — Details
(p. 349)
•
Pulse Measurements — Instructions
(p. 553)
Read More
Review the
PULSE COUNTERS
(p. 349)
and Pulse on C Terminals
sections in
CR1000 Specifications
(p. 97).
Review pulse measurement programming
in
CRBasic Editor
Help
for the
PulseCount()
and
TimerIO()
instructions.
Note
Peripheral devices are available from Campbell Scientific to expand the
number of pulse-input channels measured by the CR1000. Refer to the appendix
Measurement and Control Peripherals Lists
(p. 366)
for more information.
The figure
Pulse-Sensor Output-Signal Types
(p. 69)
illustrates pulse signal types
measurable by the CR1000:
•
low-level ac
•
high-frequency
•
switch-closure
The figure
Switch-Closure Schematic
(p. 350)
illustrates the basic internal circuit
and the external connections of a switch-closure pulse sensor. The table
Pulse
Measurements: Terminals and Programming
(p. 351)
summarizes available
measurements, terminals available for those measurements, and the CRBasic
instructions used. The number of terminals configurable for pulse input is
determined from the table
CR1000 Terminal Definitions
(p. 76).
349
Summary of Contents for CR1000
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 40: ......
Page 60: ...Section 4 System Quickstart Figure 16 PC200W View Line Graph 60 ...
Page 96: ......
Page 98: ...98 ...
Page 302: ......
Page 453: ...Section 8 Operation Figure 115 Using the Keyboard Display 453 ...
Page 456: ...Section 8 Operation Figure 118 Real Time Custom 456 ...
Page 457: ...Section 8 Operation 8 8 1 3 Final Memory Tables Figure 119 Final Memory Tables 457 ...
Page 458: ...Section 8 Operation 8 8 2 Run Stop Program Figure 120 Run Stop Program 458 ...
Page 460: ...Section 8 Operation Figure 122 File Edit 460 ...
Page 461: ...Section 8 Operation 8 8 4 PCCard Memory Card Display Figure 123 PCCard CF Card Display 461 ...
Page 478: ......
Page 506: ......
Page 536: ......
Page 636: ......
Page 642: ......
Page 644: ......
Page 676: ......
Page 677: ......