151
Section 18. Troubleshooting
Note
If any component needs to be returned to the factory for repair or
recalibration, remember that an RMA number is required. Contact a Campbell
Scientific applications engineer to receive the RMA number.
18.1 Programming
A properly deployed CR200(X) measures sensors accurately and stores all data
as instructed by its program. Experienced users analyze data soon after
deployment to ensure the CR200(X) is measuring and storing data as intended.
Most measurement and data storage problems are a result of one or more
instances of improper program code or "bugs."
18.1.1.1 Status Table as Debug Resource
Consult the CR200(X) Status Table when developing a program or when a
problem with a program is suspected.
Read More!
See
APPENDIX. Status Table and Settings
p. 15 for a complete list
of Status Table registers and hints on using the Status Table.
18.1.1.2 SkippedScan
Skipped scans are caused by long programs with short scan intervals or when
other operations occupy the processor at a scan's start time. Occasional skipped
scans may be acceptable but should be avoided. Skipped scans may compromise
frequency measurements made with pulse channels. The error occurs because
counts from a scan and subsequent skipped scans are regarded by the CR200(X)
as having occurred during a single scan. The measured frequency can be much
higher than actual. Be careful that scans that store data are not skipped. If any
scan skips repeatedly, optimization of the datalogger program or reduction of
online processing may be necessary.
18.1.1.3 VarOutOfBounds
Indicates the number of variables that are out of bounds, meaning the
dimensioned array is not large enough to store the variable. When this value is
non-zero, edit the program and increase the array size.
18.1.1.4 WatchdogErrors
Non-zero indicates the CR200(X) has crashed, which can be caused by power or
transient voltage problems, or an operating system or hardware problem.
Watchdog errors may cause telecommunications disruptions, which can make
diagnosis and remediation difficult. Sometimes a TrapCode will accompany a
WatchDogCnt.
Summary of Contents for CR200
Page 32: ...Section 2 Quickstart Tutorial 22 ...
Page 45: ...Section 3 Overview 35 3 3 Specifications ...
Page 46: ...Section 3 Overview 36 ...
Page 58: ...Section 4 Sensor Support 48 ...
Page 62: ...Section 5 Measurement and Control Peripherals 52 ...
Page 78: ...Section 8 CR200 X Configuration 68 ...
Page 102: ...Section 9 Programming 92 ...
Page 138: ...Section 11 Programming Resource Library 128 ...
Page 156: ...Section 16 Support Software 146 ...
Page 160: ...Section 17 Care and Maintenance 150 ...
Page 167: ...Section 18 Troubleshooting 157 18 3 3 2 Charging Circuit Test Solar Panel ...
Page 168: ...Section 18 Troubleshooting 158 18 3 3 3 Charging Circuit Test Transformer ...
Page 169: ...Section 18 Troubleshooting 159 18 3 3 4 Adjusting Charging Circuit Voltage ...
Page 170: ...Section 18 Troubleshooting 160 ...
Page 184: ...Appendix A Glossary 14 ...
Page 190: ...Appendix B Status Table and Settings 20 ...
Page 192: ...Appendix C Serial Port Pin Outs 22 ...
Page 211: ......