Section 3. Overview
33
3.1.8.2
Protection from Voltage Transients
Read More!
See
Grounding
(p. 55).
The CR200(X) must be grounded to minimize the risk of damage by voltage
transients associated with power surges and lightning induced transients. Earth
grounding is required to form a complete circuit for voltage clamping devices
internal to the CR200(X). Refer to
APPENDIX. Voltage Transient Suppressors
(Appendix p. 34) for information on available surge protection devices.
3.1.8.3
Calibration
Read More!
See
Self-Calibration
p. 41.
The CR200(X) uses an internal voltage reference to routinely calibrate itself. To
maintain electrical specifications, Campbell Scientific recommends factory
recalibration every two years. For calibration services, contact Campbell
Scientific to obtain a Return Materials Authorization (RMA) prior to shipping.
3.1.8.4
Replacing the Internal Battery
Caution
Fire, explosion, and severe burn hazard! Misuse or improper
installation of the lithium battery can cause severe injury. Do not
recharge, disassemble, heat above 100°C (212°F), solder directly to the
cell, incinerate, or expose contents to water. Dispose of spent lithium
batteries properly.
The CR200(X) contains a lithium battery that operates the clock and SRAM
when the CR200(X) is not powered. The CR200(X) does not draw power from
the lithium battery while it is powered by an external source. In a CR200(X)
stored at room temperature, the lithium battery should last approximately 5
years (less at temperature extremes). Where the CR200(X) is powered most or
all of the time the lithium cell should last much longer.
The internal lithium battery must be replaced periodically. Factory replacement
is recommended. Contact Campbell Scientific to obtain an RMA prior to
shipping the CR200(X).
If the lithium cell is removed or allowed to discharge below the safe level, the
CR200(X) will still operate correctly while powered. Without the lithium
battery, the clock will reset and data is lost when power is removed.
A replacement lithium battery can be purchased from Campbell (part number
15598).
TABLE. CR200(X) Lithium Battery Specifications
p. 33 lists the
specifications of the battery. However, Campbell Scientific recommends that the
battery be replaced at the factory.
Summary of Contents for CR200
Page 32: ...Section 2 Quickstart Tutorial 22 ...
Page 45: ...Section 3 Overview 35 3 3 Specifications ...
Page 46: ...Section 3 Overview 36 ...
Page 58: ...Section 4 Sensor Support 48 ...
Page 62: ...Section 5 Measurement and Control Peripherals 52 ...
Page 78: ...Section 8 CR200 X Configuration 68 ...
Page 102: ...Section 9 Programming 92 ...
Page 138: ...Section 11 Programming Resource Library 128 ...
Page 156: ...Section 16 Support Software 146 ...
Page 160: ...Section 17 Care and Maintenance 150 ...
Page 167: ...Section 18 Troubleshooting 157 18 3 3 2 Charging Circuit Test Solar Panel ...
Page 168: ...Section 18 Troubleshooting 158 18 3 3 3 Charging Circuit Test Transformer ...
Page 169: ...Section 18 Troubleshooting 159 18 3 3 4 Adjusting Charging Circuit Voltage ...
Page 170: ...Section 18 Troubleshooting 160 ...
Page 184: ...Appendix A Glossary 14 ...
Page 190: ...Appendix B Status Table and Settings 20 ...
Page 192: ...Appendix C Serial Port Pin Outs 22 ...
Page 211: ......