Section 4. Sensor Support
41
4.2.5
Self-Calibration
A calibration measurement to measure the ground offset is made at the
beginning of each measurement instruction that includes a voltage
measurement. This calibration takes about 400 microseconds. Only one
calibration measurement is made per instruction regardless of the number of
reps.
The battery voltage is checked every 8 seconds to ensure it is within the
allowable range.
4.3 Bridge Resistance Measurements
Many sensors detect phenomena by way of change in a resistive circuit.
Thermistors, strain gages, and position potentiometers are examples. Resistance
measurements are special case voltage measurements. By supplying a precise,
known voltage to a resistive circuit, then measuring the returning voltage,
resistance can be calculated.
Two bridge measurement instructions are included in the CR200(X), ExDelSE
() and Therm109 (). ExDelSE () is used with sensors that have a simple half
bridge circuit. Therm109 () is used with Campbell Scientific’s 109-L thermistor
probe. Sensors with bridge circuits that require a differential voltage
measurement, such as full bridge or 3-wire half bridge, cannot be measured with
the CR200(X).
FIGURE. Half Bridge Circuit Used with ExDelSE
(p. 41) shows the circuit that
is typically measured with ExDelSE (). In the diagram. Rs is normally the
sensor and Rf is normally a precision fixed (static) resistor. Vx is the excitation
voltage (either 2500 or 5000 mV) and V1 is the voltage (mV) measured by the
analog input channel.
Calculating the resistance of a sensor that is one of the legs of a resistive bridge
requires additional processing following the bridge measurement instruction.
FIGURE. Half Bridge Circuit Used with ExDelSE
(p. 41) lists the schematics of
a typical half bridge configuration and the calculations necessary to compute the
resistance of any single resistor, provided the value of the other resistor in the
bridge circuit is known.
Figure 28: Voltage Excitation Bridge Circuit
Summary of Contents for CR200
Page 32: ...Section 2 Quickstart Tutorial 22 ...
Page 45: ...Section 3 Overview 35 3 3 Specifications ...
Page 46: ...Section 3 Overview 36 ...
Page 58: ...Section 4 Sensor Support 48 ...
Page 62: ...Section 5 Measurement and Control Peripherals 52 ...
Page 78: ...Section 8 CR200 X Configuration 68 ...
Page 102: ...Section 9 Programming 92 ...
Page 138: ...Section 11 Programming Resource Library 128 ...
Page 156: ...Section 16 Support Software 146 ...
Page 160: ...Section 17 Care and Maintenance 150 ...
Page 167: ...Section 18 Troubleshooting 157 18 3 3 2 Charging Circuit Test Solar Panel ...
Page 168: ...Section 18 Troubleshooting 158 18 3 3 3 Charging Circuit Test Transformer ...
Page 169: ...Section 18 Troubleshooting 159 18 3 3 4 Adjusting Charging Circuit Voltage ...
Page 170: ...Section 18 Troubleshooting 160 ...
Page 184: ...Appendix A Glossary 14 ...
Page 190: ...Appendix B Status Table and Settings 20 ...
Page 192: ...Appendix C Serial Port Pin Outs 22 ...
Page 211: ......