Section 7. Installation
243
'PERFORM A MULTIPLIER CALIBRATION.
'Start by setting variable CalMode = 1. Finished when variable CalMode = 6.
'FieldCal(Function, MeasureVar, Reps, MultVar, OffsetVar, Mode, KnownVar, Index, Avg)
FieldCal
(3,WC,1,WCMultiplier,0,CalMode,KnownWC,1,30)
'If there was a calibration, store it into data table CalHist
CallTable
(CalHist)
NextScan
EndProg
7.7.12.5.5
FieldCal() Zero Basis (Opt 4) Example
Zero-basis calibration (FieldCal() instruction Option 4) is designed for use in
static vibrating wire measurements. For more information, refer to these
manuals available at www.campbellsci.com:
AVW200-Series Two-Channel VSPECT Vibrating Wire Measurement Device
CR6 Measurement and Control Datalogger Operators Manual
7.7.12.6 Field Calibration Strain Examples
Related Topics:
• Strain Measurements — Overview
(p. 74)
• Strain Measurements — Details
(p. 378)
• FieldCalStrain() Examples
(p. 243)
Strain-gage systems consist of one or more strain gages, a resistive bridge in
which the gage resides, and a measurement device such as the CR3000
datalogger. The FieldCalStrain() instruction facilitates shunt calibration of
strain-gage systems and is designed exclusively for strain applications wherein
microstrain is the unit of measure. The FieldCal() instruction (see Field
Calibration Examples
(p. 232))
is typically used in non-microstrain applications.
Shunt calibration of strain-gage systems is common practice. However, the
technique provides many opportunities for misapplication and misinterpretation.
This section is not intended to be a primer on shunt-calibration theory, but only to
introduce use of the technique with the CR3000 datalogger. Campbell Scientific
strongly urges users to study shunt-calibration theory from other sources. A
thorough treatment of strain gages and shunt-calibration theory is available from
Vishay using search terms such as 'micro-measurements', 'stress analysis', 'strain
gages', 'calculator list', at:
http://www.vishaypg.com
7.7.12.6.1
FieldCalStrain() Shunt Calibration Concepts
1. Shunt calibration does not calibrate the strain gage itself.
2. Shunt calibration does compensate for long leads and non-linearity in the
resistive bridge. Long leads reduce sensitivity because of voltage drop.
FieldCalStrain() uses the known value of the shunt resistor to adjust the gain
(multiplier / span) to compensate. The gain adjustment (S) is incorporated by
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......