Section 7. Installation
256
To enter the SDI-12 transparent mode, enter the datalogger support software
terminal emulator as shown in the figure Entering SDI-12 Transparent Mode
(p.
Press Enter until the CR3000 responds with the prompt CR3000>. Type
SDI12 at the prompt and press Enter. In response, the query Enter Cx Port is
presented with a list of available ports. Enter the port number assigned to the
terminal to which the SDI-12 sensor is connected. For example, port 1 is entered
for terminal C1. An Entering SDI12 Terminal response indicates that SDI-12
transparent mode is active and ready to transmit SDI-12 commands and display
responses.
FIGURE 66: Entering SDI-12 Transparent Mode
7.7.15.1.1
SDI-12 Transparent Mode Commands
Commands have three components:
•
Sensor address (a) — a single character, and is the first character of the
command. Sensors are usually assigned a default address of zero by the
manufacturer. Wildcard address (?) is used in the Address Query
command. Some manufacturers may allow it to be used in other
commands.
•
Command body (for example, M1) — an upper case letter (the
“command”) followed by alphanumeric qualifiers.
•
Command termination (!) — an exclamation mark.
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......