FIGURE 71: PT100 Resistance() Basic-Circuit Series Schematic
PT100 Resistance() Basic-Circuit Measurement
'This program example demonstrates the measurement of 100-ohm PRT (PT100) with
'current excitation. See previous procedure and schematic.
'Declare constants and variables:
Const
RS0 = 100000
'Resistance of PT100 at 0 °C from calibration program
Public
X(3)
'Raw output from the bridge
Public
RS(3)
'Calculated PT100 resistance
Public
RS_RS0(3)
'Calculated ratio RS/RS0
Public
DegC(3)
'Calculated temperature
Public
Cnt
'Counter for For/Next loop
BeginProg
Scan
(1,Sec,0,0)
....'Measure X:
'Resistance(Dest,Reps,Range,InputChan,IExChan,MPE,ExuA,RevEx,RevDiff,
' SettlingTime,
Integ
,Mult,Offset)
Resistance
(X(),3,mV50,1,Ix1,3,1000,True,True,0,250,1,0)
'Calculate RS and RS_RS0:
For
Cnt = 1
To
3
RS(Cnt) = X(Cnt)
RS_RS0(Cnt) = RS(Cnt) / RS0
Next
Cnt
Cnt = 0
....'Calculate temperature from RS_RS0:
'PRTCalc(Dest,Reps,Source,PRTType,Mult,Offset)
PRTCalc
(DegC(),3,RS_RS0(),1,1.0,0)
NextScan
EndProg
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......