Section 8. Operation
374
21 segments. So, (21 segments) • (4 s / seg
ment) = 84 s per complete auto self
-
calibration. The worst-
case is (91 segments) • (4 s / segment) = 364 s per
complete auto self-calibration.
During instrument power-up, the CR3000 computes calibration coefficients by
averaging ten complete sets of auto self-calibration measurements. After power
up, newly determined G and B values are low-pass filtered as follows:
Next_Value = (1/5) • (new value) + (4/5) • (old value)
This results in the following settling percentages:
•
20% for 1 new value,
•
49% for 3 new values
•
67% for 5 new values
•
89% for 10 new values
•
96% for 14 new values
If this rate of update is too slow, the Calibrate() instruction can be used. The
Calibrate() instruction computes the necessary G and B values every scan
without any low-pass filtering.
For a VoltSe() instruction, B is determined as part of auto self-calibration only if
the parameter MeasOff
=
0. An exception is B for VoltSe() on the
±
2500 input
range with a 250
µ
s integration, which is always determined in auto self-
calibration for use internally. For a VoltDiff() instruction, B is determined as part
of auto self-calibration only if the parameter RevDiff
=
0.
VoltSe() and VoltDiff() instructions, on a given input range with the same
integration durations, use the same G values but different B values. The five
input-voltage ranges (±5000 mV, ±1000 mV, ±200 mV, and ±50 mV), in
combination with the three most common integration durations (250
µ
s, 50 Hz
half-cycle, and 60 Hz half-cycle) result in a maximum of 15 different gains (G),
and 15 offsets for VoltSe() measurements (B), and 15 offsets for VoltDiff()
measurements (B) to be determined during auto self-calibration (maximum of 45
values). These values can be viewed in the Status table, with entries identified as
listed in table CalGain() Field Descriptions
(p. 375)
Auto self-calibration can be overridden with the Calibrate() instruction, which
forces a calibration for each execution, and does not employ low-pass filtering on
the newly determined G and B values. The Calibrate() instruction has two
parameters: CalRange and Dest. CalRange determines whether to calibrate only
the necessary input ranges for a given CRBasic program (CalRange
= 0) or to
calibrate all input ranges (CalRange
≠ 0). The
Dest parameter should be of
sufficient dimension for all returned G and B values, which is a minimum of two
for the auto self-calibration of VoltSE() including B (offset) for the ±5000 mV
input range with first 250 µs integration, and a maximum of 45 for all input-
voltage ranges used and possible integration durations.
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......