Section 10. Troubleshooting
519
Battery Test
If using a rechargeable power supply,
disconnect the charging source (i.e., solar panel
or ac transformer) from the battery pack. Wait
20 minutes before proceeding with this test.
Test Voltage at Charging Regulator
Set a voltmeter to read dc voltage as high as 15
V. Measure the voltage between a 12V and G
terminal on the charging regulator.
Is the voltage > 11.0 Vdc?
No
Yes
Test the Battery Under Load
Program the CR3000 to measure battery
voltage using a 0.01-second scan rate. Use the
voltmeter to measure the voltage between a
12V and G terminal on the charging regulator.
Is the voltage > 10.8 Vdc?
Is the battery a sealed, rechargeable
battery?
No
No
Yes
Is the voltage ≥ 10.5 Vdc?
Replace battery / batteries*
No
Yes
Yes
Recharge battery*
Is the battery voltage > 12 Vdc?
Battery voltage is adequate for CR3000 operation. However, if the CR3000 is to function for a long
period, Campbell Scientific recommends replacing, or, if using a sealed, rechargeable battery,
recharging the battery so the voltage is > 12 Vdc.
No
Yes
The battery is good.
*When using a sealed, rechargeable battery that is recharged with primary power provided by solar panel or ac/ac - ac/dc transformer, testing the charging regulator
is recommended. See Charging Regulator with Solar Panel Test
(p. 519)
or Charging Regulator with Transformer Test
(p. 521).
10.9.3.2 Charging Regulator with Solar Panel Test
The procedure outlined in this flow chart tests PS100 and CH100 charging
regulators that use solar panels as the power source. If a need for repair is
indicated after following the procedure, see Assistance
(p. 5)
for information on
sending items to Campbell Scientific.
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......