Section 7. Installation
109
Note Lightning strikes may damage or destroy the CR3000 and
associated sensors and power supplies.
In addition to protections discussed in
,
use of a simple lightning rod and low-
resistance path to earth ground is adequate protection in many installations. .
FIGURE 33: Lightning Protection Scheme
7.3.2 Single-Ended Measurement Reference
Low-level, single-ended voltage measurements (<200 mV) are sensitive to ground
potential fluctuation due to changing return currents from 12V, SW12, 5V, and
C1 – C8 terminals. The CR3000 grounding scheme is designed to minimize