Section 7. Installation
235
'DECLARE VARIABLE FOR FieldCal() CONTROL
Public
CalMode
'DECLARE DATA TABLE FOR RETRIEVABLE CALIBRATION RESULTS
DataTable
(CalHist,NewFieldCal,200)
SampleFieldCal
EndTable
BeginProg
'LOAD CALIBRATION CONSTANTS FROM FILE CPU:CALHIST.CAL
'Effective after the zero calibration procedure (when variable CalMode = 6)
LoadFieldCal
(true)
Scan
(100,mSec,0,0)
'SIMULATE SIGNAL THEN MAKE THE MEASUREMENT
'Zero calibration is applied when variable CalMode = 6
ExciteV
(Vx1,SimulatedRHSignal,0)
VoltSE
(RH,1,mV2000,1,1,0,250,0.05,RHOffset)
'PERFORM A ZERO CALIBRATION.
'Start by setting variable CalMode = 1. Finished when variable CalMode = 6.
'FieldCal(Function, MeasureVar, Reps, MultVar, OffsetVar, Mode, KnownVar, Index, Avg)
FieldCal
(0,RH,1,0,RHOffset,CalMode,KnownRH,1,30)
'If there was a calibration, store calibration values into data table CalHist
CallTable
(CalHist)
NextScan
EndProg
7.7.12.5.2
FieldCal() Offset (Opt 1) Example
Most CRBasic measurement instructions have a multiplier and offset parameter.
FieldCal() Option 1 adjusts the offset argument such that the output of the sensor
being calibrated is set to the magnitude of the FieldCal() KnownVar parameter.
Subsequent measurements have the same offset added. Option 0 does not affect
the multiplier argument. Option 0 does not affect the multiplier argument.
Example Case: A sensor measures the salinity of water. Multiplier is known to be
stable, but sensor offset drifts and requires regular offset correction using a
standard solution. The following procedure offsets the measurement to obtain the
calibration report shown.
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......