Section 7. Installation
247
FIGURE 63: Strain Gage Shunt
Calibration Finish
7.7.12.6.4
FieldCalStrain() Quarter-Bridge Zero
Continuing from FieldCalStrain() Quarter-Bridge Shunt Example
(p. 246),
keep the
249 kΩ resistor in place to simulate a strain. Using the CR1000KD
Keyboard/Display or software numeric monitor, change the value in variable
Zero_Mode to 1 to start the zero calibration as shown in figure Zero Procedure
Start
(p. 247).
When Zero_Mode increments to 6, zero calibration is complete as
shown in figure Zero Procedure Finish
(p. 247).
FIGURE 64: Zero Procedure Start
FIGURE 65: Zero Procedure Finish