Section 7. Installation
301
PRTType codes depend on the alpha value of the PRT, which is determined and
published by the PRT manufacturer.
PRTCalc() PRTType = 1
, α = 0.00385
1
Constants
Coefficient
a
3.9083000E-03
d
-2.3100000E-06
e
1.7584810E-05
f
-1.1550000E-06
g
1.700
h
-2.900
i
9.100
j
2.502
1
Compliant with the following standards: IEC 60751:2008 (IEC 751), ASTM
E1137-04, JIS 1604:1997, EN 60751, DIN43760, BS1904, and others
(reference IEC 60751 and ASTM E1137), α = 0.00385
PRTCalc() PRTType = 2
, α = 0.00392
1
Constant
Coefficient
a
3.9786300E-03
d
-2.3452400E-06
e
1.8174740E-05
f
-1.1726200E-06
g
1.700
h
-2.700
i
8.800
j
2.502
1
US Industrial Standard, α = 0.00392 (Reference: Logan Enterprises)