Section 8. Operation
411
closure frequency is less than the maximum high-frequency measurement
frequency. Sensors that commonly output a switch closure or open-collector
signal include:
•
Tipping-bucket rain gages
•
Switch closure anemometers
•
Flow meters
Data output options include counts, frequency (Hz), and running average.
P Terminals
An internal 100 k
Ω
pull-up resistor pulls an input to 5 Vdc with the switch open,
whereas a switch closure to ground pulls the input to 0 V. An internal hardware
debounce filter has a 3.3 ms time-constant. Connection configurations are
illustrated in table
.
•
Maximum input f
requency = 90 Hz
CRBasic instruction: PulseCount()
An internal 100 k
Ω
pull-up resistor pulls an input to 5 Vdc with the switch open,
whereas a switch closure to ground pulls the input to 0 V. An internal hardware
debounce filter has a 3.3 ms time-constant. Connection configurations are
illustrated in table
.
•
Maximum input frequency = 90 Hz
•
CRBasic instruction: PulseCount()
C Terminals
Switch closure mode is a special case edge-count function that measures dry-
contact-switch closures or open collectors. The operating system filters bounces.
Connection configurations are illustrated in table Switch Closures and Open
Collectors
(p. 413).
•
Maximum input frequency = 150 Hz
•
CRBasic instruction: PulseCount()
8.1.3.5 Edge Timing
Edge time and period can be measured on P or C terminals. Applications for edge
timing include:
•
Measurements for feedback control using pulse-width or pulse-duration
modulation (PWM/PDM).
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......