95
6. Specifications
1.1 -- 8 10 30
CR3000 specifications are valid from ─25° to 50°C in non
-condensing environments unless otherwise specified. Recalibration is recommended every three years. Critical specifications and system
configurations should be confirmed with a Campbell Scientific sales engineer before purchase.
2.0 -- 8 10 30
PROGRAM EXECUTION RATE
2.1 -- 8 10 30
10 ms to one day at 10 ms increments
3.0 -- 8 10 30
ANALOG INPUTS (SE 1–
28, DIFF 1
–14)
3.0.1 -- 8 10 30
Fourteen differential (DIFF) or 28 single-ended (SE) individually
configured input channels. Channel expansion provided by
optional analog multiplexers.
3.1.0 -- 8 10 30
RANGES and RESOLUTION: With reference to the following table,
basic resolution (Basic Res) is the resolution of a single
A/D
conversion. A DIFF measurement with input reversal has better
(finer) resolution by twice than Basic Res.
3.1.1 -- 30
Range (mV)
1
DIFF Res, μV
2
(17 bit)
Basic Res, μV
(16 bit)
±5000
±1000
±200
±50
±20
83.33
16.67
3.33
0.83
0.33
167
33.3
6.67
1.67
0.67
1
Range overhead of ≈9% on all ranges guarantees full
-scale
voltage will not cause over-range.
2
Resolution of DIFF measurements with input reversal.
3.2 -- 30
ANALOG INPUT ACCURACY
3
:
±(0.04% of r offset), 0° to 40°C
±(0.07% of r offset),
-
25° to 50°C
±(0.09% of r offset),
-
40° to 85°C (
-XT only)
3.2.1 -- 8 10 30
3
Accuracy does not include sensor and measurement noise.
Offset definitions:
Offset = 1.5 x Basic Res + 1.0 µV (for DIFF measurement w/ input
reversal)
Offset = 3 x Basic Res + 2.0 µV (for DIFF measurement w/o input
reversal)
Offset = 3 x Basic Res + 5.0 µV (for SE measurement)
3.3 -- 8 10 30
ANALOG MEASUREMENT SPEED:
3.3.1 -- 30
---Total Time
4
---
Inte-
gration
Type
Code
Inte-
gration
Time
Settling
Time
SE
with
no
Rev
DIFF
with
Input
Rev
250
_60Hz
5
_50Hz
5
250 µs
16.67 ms
20.00 ms
200 µs
3 ms
3 ms
≈0.7 ms
≈20 ms
≈23 ms
≈1.4 ms
≈40
ms
≈46 ms
4
Includes 250 μs for conversion to engineering units.
5
AC line noise filter
3.4 -- 8 10 30
3.4.1 -- 8 10 30
INPUT-NOISE VOLTAGE: For DIFF measurements with input
reversal on ±20 mV input range (digital resolution dominates for
higher ranges):
250 μs Integration: 0.4 μV RMS
50/60 Hz Integration: 0.19 μV RMS
3.4.2 -- 8 10 30
INPUT LIMITS: ±5 Vdc
3.4.3 -- 8 10 30
DC COMMON-MODE REJECTION: >100 dB
3.4.4 -- 8 10 30
NORMAL-
MODE REJECTION: 70 dB @ 60 Hz when using 60 Hz
rejection
3.4.5 -- 8 10 30
INPU
T VOLTAGE RANGE W/O MEASUREMENT CORRUPTION: ±8.6
Vdc max.
3.4.6 -- 8 10 30
SUSTAINED-
INPUT VOLTAGE W/O DAMAGE: ±16 Vdc max
3.4.7 -- 8 10 30
INPUT CURRENT: ±1 nA typical, ±6 nA max. @ 50°C; ±120 nA @
85°C
3.4.8 -- 8 10 30
INPUT RESISTANCE: 20 GΩ typical
3.4.9 -- 8 10 30
ACCURACY OF BUILT-
IN REFERENCE JUNCTION THERMISTOR (for
thermocouple measurements):
±0.3°C,
-
25° to 50°C
±0.8°C,
-
55° to 85°C (
-XT only)
4.0 -- 8 10 30
ANALOG OUTPUTS (VX 1–
4, IX 1
–
3, CAO 1
–
2)
4.0.1 -- 30
Four switched voltage and three switched current outputs
sequentially active only during measurement. Two continuous
outputs.
4.0.2 -- 8 10 30
RANGES AND RESOLUTION:
4.1 -- 30
Chan-
nel
Range
Res-
olution
Current
Source
/ Sink
Compli-
ance
Voltage
VX 1–4
IX 1–3
CAO 1–2
±5 Vdc
±2.5 mA
±5 Vdc
0.17 mV
0.08 µA
0.17 mV
±50 mA
N/A
±15 mA
N/A
±5 Vdc
N/A
4.2 -- 30
ANALOG OUTPUT ACCURACY (VX and CAO):
±(0.04% of s 0.5 mV, 0° to 40°C
±(0.07% of s 0.5 mV,
-
25° to 50°C
±(0.09% of s 0.5 mV,
-
40° to 85°C (
-XT only)
4.3 -- 30
ANALOG OUTPUT ACCURACY (IX)
±(0.1% of s 0.5 µA, 0° to 40°C
±(0.13% of s 0.5 µA,
-
25° to 50°C
±(0.15% of s 0.5 µA,
-
40° to 85°C (
-XT only)
4.4 -- 8 10 30
3.5.0 -- 8 10 30
VX FREQUENCY SWEEP FUNCTION: Switched outputs provide a
programmable swept frequency, 0 to 5000 mV square waves for
exciting vibrating wire transducers.
PERIOD AVERAGE
3.5.0a -- 8 10 30
Any of the 28 SE analog inputs can be used for period averaging.
Accuracy is ±(0.01% of r resolution), where resolut
ion is
68 ns divided by the specified number of cycles to be measured.
INPUT AMPLITUDE AND FREQUENCY:
3.5.1 -- 30
Volt-
Input
Signal
Peak
-
Peak
Min
Pulse
Max
age
Gain
Range
Code
Min
mV
6
Max
V
7
Width
µs
Freq
kHz
8
1
5
20
50
mV1000
mV200
mV50
mV20
200
20
5
2
10
2
2
2
2.5
5.0
10.0
100
200
100
50
20
6
Signal to be centered around
Threshold
(see
PeriodAvg()
instruction).
7
Signal to be centered around ground.
8
The maximum frequency = 1/(twice minimum pulse width)
for 50% of duty cycle signals.
5.0 -- 8 10 30
RATIOMETRIC MEASUREMENTS
5.1 -- 30
MEASUREMENT TYPES: The CR3000 provides ratiometric
resistance measurements using voltage or current excitation. Four
switched voltage excitation outputs are available for
measurement of four- and six-wire full bridges, and two-, three-,
and four-wire half bridges. Three switched current excitation
outputs are available for direct resistance measurements.
Optional excitation polarity reversal minimizes dc errors.
5.2 -- 30
RATIOMETRIC MEASUREMENT ACCURACY
9,10,11
:
±(0.0
2% of Voltage Measu Offset
12
), 0° to 40°C
±(0.025% of Voltage Measu Offset
12
), ─25° to 50°C
±(0.03% of Voltage Measu Offset
12
), ─40° to 85°C (
-XT)
5.2.1 -- 8 10 30
9
Accuracy specification assumes excitation reversal for excitation
v
oltages < 1000500 mV and excitation currents < 500 µA.
Assumption does not include bridge resistor errors and sensor
and measurement noise.
10
For
Resistance()
instruction, the sensor resistance is determined
from VS
/ I
X, where excitation current IX is measured across a
1000 Ω, ±0.005% at 25˚C, 2 ppm•˚C
-1
TCR internal resistor.
11
Estimated accuracy, ∆X (where X is value returned from
measurement with
Multiplier
=1,
Offset
= 0):
BRHalf()
Instruction: ∆X = ∆V1/
VX.
BRFull()
Instruction:
∆X = 1000 x ∆V1/
VX
, expressed as mV•V
-1
.
Note
∆V
1 is calculated from the ratiometric measurement
accuracy. See manual section
Resistance Measurements
for
more information.
12
Offset definitions:
Offset = 1.5 x Basic Res + 1.0 µV (for DIFF measurement w/ input
reversal)
O
ffset = 3 x Basic Res + 2.0 µV (for DIFF measurement w/o input
reversal)
Offset = 3 x Basic Res + 5.0 µV (for SE measurement)
Note
Excitation reversal reduces offsets by a factor of two.
6.0 -- 8 10 30
PULSE COUNTERS (P 1–4)
6.0.1 -- 8 10 30
Four inputs individually selectable for switch closure, high-
frequency pulse, or low-level ac. Independent 24-bit counters for
each input.
6.1 -- 8 10 30
MAXIMUM COUNTS PER SCAN: 16.8 x 10
6
6.2 -- 8 10 30
SWITCH CLOSURE MODE:
Minimum Switch Closed Time: 5 ms
Minimum Switch Open Time: 6 ms
Max. Bounce Time: 1 ms open without being counted
6.3 -- 8 10 30
HIGH
-FREQUENCY PULSE MODE:
Maximum-
Input Frequency: 250 kHz
Maximum-Input Voltage: ±20 V
Voltage Thresholds: Count upon transition from below 0.9 V to
above 2.2 V after
input filter with 1.2 μs time constant.
6.4 -- 8 10 30
LOW-LEVEL AC MODE: Internal ac coupling removes dc offsets up
to ±0.5 Vdc.
Input Hysteresis: 12 mV RMS @ 1 Hz
Maximum ac-Input Voltage: ±20 V
Minimum ac-Input Voltage:
6.4.1 -- 8 10 30
Sine wave (mV RMS)
Range (Hz)
20
200
2000
5000
1.0 to 20
0.5 to 200
0.3 to 10,000
0.3 to 20,000
7.0 -- 8 10 30
DIGITAL I/O PORTS (C 1
–
8, SDM 1
–3)
7.0.1 -- 8 10 30
Eight ports software selectable as binary inputs or control outputs.
Provide on/off, pulse width modulation, edge timing, subroutine
interrupts / wake up, switch closure pulse counting, high
-
frequency pulse counting, asynchronous communications (UARTs),
and SDI-12 communications.
7.1 -- 8 10 30
LOW FREQUENCY MODE MAX: <1 kHz
7.2 -- 8 10 30
HIGH FREQUENCY MODE MAX: 400 kHz
7.3 -- 8 10 30
SWITCH
-
CLOSURE FREQUENCY MAX: 150 Hz
7.4 -- 8 10 30
EDGE-TIMING RESOLUTION: 136 ns
7.5 -- 8 10 30
OUTPUT VOLTAGES (no
load): high 5.0 V ±0.1 V; low < 0.1 V
7.6 -- 8 10 30
OUTPUT RESISTANCE: 330 Ω
7.7 -- 8 10 30
INPUT STATE: high 3.8 to 16 V; low
-8.0 to 1.2 V
7.8 -- 8 10 30
INPUT HYSTERISIS: 1.4 V
7.9 -- 8 10 30
INPUT RESISTANCE:
100 kΩ with inputs < 6.2 Vdc
220 Ω with inputs ≥ 6.2 Vdc
7.10 -- 8 10 30
SERIAL DEVICE / RS
-232 SUPPORT: 0 to 5 Vdc UART
7.11 -- 30
ADDITIONAL DIGITAL PORTS: SDM-C1, SDM-C2, SDM-C3 are
dedicated for measuring SDM devices.
7.12 -- 8 10 30
SWITCHED 12 Vdc (SW12V)
Two independent 12 Vdc unregulated terminals switched on and
off under program control. Thermal fuse hold current = 900 mA at
20°C, 650 mA at 50°C, and 360 mA at 85°C.
8.0 -- 8 10 30
COMPLIANCE
8.1 -- 8 10 30
View the EU Declaration of Conformity at
9.0 -- 8 10 30
COMMUNICATION
9.1 -- 8 10 30
RS-232 PORTS:
DCE nine-pin: (electrically isolated) for computer connection or
connection of modems not manufactured by Campbell Scientific.
COM1 to COM4: four independent Tx/Rx pairs on control ports
(non-
isolated); 0 to 5 Vdc
UART
Baud Rate: selectable from 300 bps to 115.2 kbps.
Default Format: eight data bits; one stop bits; no parity.
Optional Formats: seven data bits; two stop bits; odd, even parity.
9.2 -- 8 10 30
CS I/O PORT: Interface with comms peripherals manufactured
by
Campbell Scientific.
9.3 -- 8 10 30
SDI-12: Digital control ports C1, C3, C5, C7 are individually
configurable and meet SDI-12 Standard v. 1.3 for datalogger
mode. Up to ten SDI-12 sensors are supported per port.
9.4 -- 10 30
PERIPHERAL PORT: 40
-pin interface for attaching CompactFlash or
Ethernet peripherals.
9.5 -- 8 10 30
PROTOCOLS SUPPORTED: PakBus, AES
-
128 Encrypted PakBus,
Modbus, DNP3, FTP, HTTP, XML, HTML, POP3, SMTP, Telnet,
NTCIP, NTP, web API, SDI-12, SDM.
10.0 -- 8 10 30
SYSTEM
10.1 -- 8 10 30
PROCESSOR: Renesas H8S 2674 (16
-bit CPU with 32-bit internal
core running at ≈21.9 MHz)
10.2 -- 8 10 30
MEMORY: 2 MB of flash for operating system; 4 MB of battery
-
backed SRAM for CPU, CRBasic programs, and data.
10.3 -- 8 10 30
REAL-TIME CLOCK ACCURACY: ±3 min. per year. Correction via GPS
optional.
10.4 -- 8 10 30
RTC CLOCK RESOLUTION: 10 ms
11.0 -- 8 10 30
SYSTEM POWER REQUIREMENTS
11.1 -- 8 10 30
VOLTAGE: 10 to 16 Vdc
11.2 -- 30
INTERNAL BATTERIES: 1200-
mAhr lithium battery for clock and
SRAM backup. Typically provides 3 years of backup. Optional
10-
Ahr alkaline or 7
-
Ahr rechargeable b base available as
primary power supply.
11.3 -- 8 10 30
EXTERNAL BATTERIES: Optional 12 Vdc nominal alkaline and
rechargeable available. Power connection is reverse polarity
protected.
11.4 -- 8 10 30
TYPICAL CURRENT DRAIN at 12 Vdc:
Sleep Mode: 2 mA
1 Hz Sample Rate (one fast SE meas.): 3 mA
100 Hz Sample Rate (one fast SE meas.): 10 mA
100 Hz Sample Rate (one fast SE meas. with RS
-232
communications): 38 mA
Active integrated keyboard display adds 1 mA (42 mA with
backlight on).
12.0 -- 8 10 30
PHYSICAL
12.1
DIMENSIONS: 241 x 178 x 96 mm (9.5 x 7.0 x 3.8 in.); additional
clearance required for cables and leads.
12.2
MASS / WEIGHT:
12.2.1 -- 30
Base Type:
Mass (kg)
Weight (lbs)
Low Prof Base
Alkaline Base
Recharge Base
1.6
3.8
4.8
3.6
8.3
10.7
13.0
WARRANTY
13.1
Warranty is stated in the published price list and in opening pages
of this and other user manuals.
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......