Section 6. Specifications
106
6.1.1.3 Thermistor Measurements — Specifications
The CR6 can measure two-wire thermistors directly using an on-
board 5 kΩ
resistor to complete the bridge. A
U
terminal pair both excites and measures the
thermistor.
Input Resistance:
5 kΩ ±0.1%, 10 ppm/°C
bridge-completion resistor
Measurement
Type:
Single-ended voltage
Range:
±5000 mV
Accuracy:
±0.25% of reading
Resolution:
0.001 (Ω RMS)
Speed (vibrating wire and
thermistor combined):
< 1 s
6.1.1.4 Resistance Measurements — Specifications
The CR6 makes ratiometric-resistance measurements with voltage and current
excitation.
Resistive Bridges with Voltage
Excitation:
•
Four-wire full bridge
•
Six-wire full bridge
•
Two-wire half bridge
•
Three-wire half bridge
•
Four-wire half bridge
Resistive Bridges with Current
Excitation:
•
Four-wire full bridge
•
Four-wire resistor
Excitation Polarity
Reversal:
Minimizes dc error
Accuracy
2,3,4
0 to 40 °C:
±(0.02% of voltage measu offset
5
)
–40 to 70 °C:
±(0.025% of voltage meas offset
5
)
–55 to 85 °C:
±(0.03% of voltage meas offset
5
)
1
Typically, at least one terminal is configured for excitation output. Multiple
sensors can usually use a common excitation terminal.
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......