Section 7. Installation
282
software documentation available at
www.campbellsci.com
. Be aware of the
following precautions:
•
The CR6 does not check for out-of-bounds values in mode variables.
•
Valid mode variable entries are
1
or
4
.
Before, during, and after calibration, one of the following codes will be stored in
the
CalMode
variable:
FieldCal() Codes
Value Returned
State
-1
Error in the calibration setup
-2
Multiplier set to
0
or
NAN
; measurement =
NAN
-3
Reps
is set to a value other than
1
or the size of
MeasureVar
0
No calibration
1
Ready to calculate (
KnownVar
holds the first of a
two point calibration)
2
Working
3
First point done (only applicable for two point
calibrations)
4
Ready to calculate (
KnownVar
holds the second of a
two-point calibration)
5
Working (only applicable for two point calibrations)
6
Calibration complete
7.7.12.4.1 One-Point Calibrations (Zero or Offset)
Zero operation applies an offset of equal magnitude but opposite sign. For
example, when performing a zeroing operation on a measurement of 15.3, the
value –15.3 will be added to subsequent measurements.
Offset operation applies an offset of equal magnitude and same sign. For
example, when performing an offset operation on a measurement of 15.3, the
value 15.3 will be added to subsequent measurements.
See
FieldCal() Zero or Tare (Opt 0) Example
(p. 284)
and
FieldCal() Offset (Opt 1)
Example
(p. 286)
for demonstration programs:
1. Use a separate
FieldCal()
instruction and variables for each sensor to be
calibrated. In the CRBasic program , put the
FieldCal()
instruction
immediately below the associated measurement instruction.
2.
Set mode variable = 0 or 6 before starting.
3. Place the sensor into zeroing or offset condition.
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......