Section 8. Operation
436
Analog Voltage Measurement Offsets
Input
Voltage Range
(mV)
Differential
Measurement
With Input or
Excitation
Reversal
(
µ
V)
Differential
Measurement
Without Input
or
Excitation
Reversal
(
µ
V)
Single-Ended
(
µ
V)
±
5000
±
40
±
70
±
140
±
1000
±
10
±
30
±60
±
200
±3
±5
±10
Analog Voltage Measurement Resolution
Input
Voltage Range
(mV)
Differential
Measurement
With Input or
Excitation Reversal
f
N1
= 60
(RMS
µ
V)
Differential
Measurement
With Input or
Excitation Reversal
f
N1
= 60
(Bits)
±
5000
1.2
23
±
1000
0.24
23
±
200
0.1
22
Note
— see
Specifications
(p. 101)
for a complete tabulation of measurement
resolution
As an example, figure
Voltage Measurement Accuracy Band Example
(p. 404)
shows changes in accuracy as input voltage changes on the ±5000 input range.
Percent-of-reading is the principle component, so accuracy improves as input
voltage decreases. Offset is very small, but could be significant in applications
wherein the sensor-signal voltage is near 0 V.
Offset depends on measurement type and voltage-input range. Offsets are
tabulated in table Analog Voltage Measurement Offsets
.
For example, for a
differential measurement with input reversal on the ±5000 mV input range, the
offset voltage is 40 µV.
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......