Section 8. Operation
415
StrainCalc Example
: See
FieldCalStrain() Examples
(p. 296).
8.1.2.6 Current Measurements — Details
Related Topics:
•
Current Measurements — Overview
(p. 75)
•
Current Measurements — Details
(p. 415)
For a complete treatment of current-loop sensors (4 to 20 mA, for example),
please consult the following publications available at
www.campbellsci.com/app-notes
:
•
Current Output Transducers Measured with Campbell Scientific
Dataloggers (2MI-B)
•
CURS100 100 Ohm Current Shunt Terminal Input Module
The CR6 is equipped to make resistive-bridge measurements with current
excitation. See
Resistance Measurements — Details
for more information.
8.1.2.7 Voltage Measurements — Details
Related Topics:
•
Voltage Measurements — Specifications
(p. 102)
•
Voltage Measurements — Overview
(p. 72)
•
Voltage Measurements — Details
(p. 415)
8.1.2.7.1 Voltage Measurement Limitations
Caution
Sustained voltages in excess of
±
16 V applied to terminals
configured for analog input can temporarily corrupt all analog
measurements.
Warning
Sustained voltages in excess of
±
20 V applied to terminals
configured for analog input will damage CR6 circuitry.
Voltage Ranges
Related Topics:
•
Voltage Measurements — Specifications
(p. 102)
•
Voltage Measurements — Overview
(p. 72)
•
Voltage Measurements — Details
(p. 415)
In general, use the smallest fixed-input range that accommodates the full-scale
output of the sensor. This results in the best measurement accuracy and resolution.
The CR6 has fixed input ranges for voltage measurements and an auto-range to
automatically determine the appropriate input voltage range for a given
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......