Section 8. Operation
416
measurement. The table
Analog Voltage Input Ranges and Options
(p. 416)
lists
these input ranges and codes.
An approximate 5% range overhead exists on fixed input voltage ranges. In
other words, over-range on the
±
1000 mV input range occurs at approximately
1050 mV and –1050 mV. The CR6 indicates a measurement over-range by
returning a
NAN
(not a number) for the measurement.
Automatic Range Finding
For signals that do not fluctuate too rapidly, range argument
AutoRange
allows
the CR6 to automatically choose the voltage range.
AutoRange
makes two
measurements. The first measurement determines the range to use. It is made
with a first notch frequency (
fN1
) of 50 kHz on the ±5000 mV range. The
second measurement is made using the range determined from the first and the
same
fN1
. Both measurements use the settling time entered in the
SettlingTime
parameter. Auto-ranging optimizes resolution but takes longer than a
measurement on a fixed range because of the two-measurement sequences.
An auto-ranged measurement will return
NAN
("not a number") if the voltage
exceeds the range picked by the first measurement. To avoid problems with a
signal on the edge of a range,
AutoRange
selects the next larger range when the
signal exceeds 90% of a range.
Use auto-ranging for a signal that occasionally exceeds a particular range.
AutoRange
should not be used for rapidly fluctuating signals, particularly signals
traversing multiple voltage ranges rapidly. The possibility exists that the signal
can change ranges between the internal range check and the actual measurement.
Analog Voltage Input Ranges and Options
Range Code
Description
mV5000
1
measures voltages between ±5000 mV
mV1000
1
measures voltages between ±1000 mV
mV200
1
measures voltages between ±200 mV
AutoRange
1
datalogger determines the most suitable
range
1
Append with
C
to enable common-mode null /
open-input detect
(p. 396)
(Example:
mV5000C
)
Input Limits / Common-Mode Range
Related Topics:
•
Voltage Measurements — Specifications
(p. 102)
•
Voltage Measurements — Overview
(p. 72)
•
Voltage Measurements — Details
(p. 415)
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......