Section 8. Operation
435
Time Skew Between Measurements
Time skew between consecutive voltage measurements is a function of settling
times, A-to-D conversion, and the number entered into the
Reps
parameter of the
VoltDiff()
or
VoltSE()
instruction. A close approximation is:
time skew = settling time + A
-to-D conversion time + reps
where A-to-D conversion time equals 1/fN1 µs. If reps (repetitions) > 1 (multiple
measurements by a single instruction), no additional time is required. If reps = 1
in consecutive voltage instructions, add 15 µs per instruction.
Measurement Accuracy
Read More
For an in-depth treatment of accuracy estimates, see the
technical paper
Measurement Error Analysis
soon available at
www.campbellsci.com/app-notes.
Accuracy describes the difference between a measurement and the true value.
Many factors affect accuracy. This section discusses the affect
percent-or-reading, offset, and resolution have on the accuracy of the
measurement of an analog voltage sensor signal. Accuracy is defined as follows:
accuracy = percent
-of-r offset
where percents-of-reading are tabulated in the table
Analog Voltage Measurement
Accuracy
(p. 402),
and offsets are tabulated in the table
Analog Voltage
Measurement Offsets
(p. 402).
Note
Error discussed in this section and error-related specifications of
the CR6 do not include error introduced by the sensor or by the
transmission of the sensor signal to the CR6.
Analog Voltage Measurement Accuracy
1
0 to 40 °C
–40 to 70 °C
–55 to 85 °C
±(0.03% of r
offset)
±(0.03% of r
offset)
±(0.03% of r
offset)
1
Assumes the CR6 is within factory specifications
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......