Section 7. Installation
218
•
Offset
•
Two-point slope and offset
•
Two-point slope only
•
Zero basis (designed for use with static vibrating wire measurements)
These demonstration programs are provided as an aid in becoming familiar with
the
FieldCal()
features at a test bench without actual sensors. For the purpose of
the demonstration, sensor signals are simulated by CR800 terminals configured
for excitation. To reset tests, use the support software
File Control
(p. 498)
menu
commands to delete .cal files, and then send the demonstration program again to
the CR800. Term equivalents are as follows:
"offset" = "y
-
intercept" = "zero"
"multiplier" = "slope" = "gain"
7.7.11.5.1 FieldCal() Zero or Tare (Opt 0) Example
Most CRBasic measurement instructions have a
multiplier
and
offset
parameter.
FieldCal()
Option 0
adjusts the
offset
argument such that the output of the sensor
being calibrated is set to the value of the
FieldCal()
KnownVar
parameter, which
is set to
0
. Subsequent measurements have the same offset subtracted.
Option 0
does not affect the
multiplier
argument.
Example Case: A sensor measures the relative humidity (RH) of air. Multiplier
is known to be stable, but sensor offset drifts and requires regular zeroing in a
desiccated chamber. The following procedure zeros the RH sensor to obtain the
calibration report shown. To step through the example, use the CR1000KD
Keyboard/Display or software
numeric monitor
(p. 506)
to change variable values as
directed.
Calibration Report for Relative Humidity Sensor
CRBasic Variable
At Deployment
At 30-Day Service
SimulatedRHSignal
output
100
mV
105
mV
KnownRH
(desiccated
chamber)
0
%
0
%
RHMultiplier
0.05
% / mV
0.05
% / mV
RHOffset
-5
%
-5.25
%
RH
0
%
0
%
1. Send CRBasic example
FieldCal() Zero
(p. 219)
to the CR800. A terminal
configured for excitation has been programmed to simulate a sensor output.
2. To place the simulated RH sensor in a simulated-calibration condition (in the
field it would be placed in a desiccated chamber), place a jumper wire between
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......