Section 7. Installation
224
a. For the first point, set variable
SimulatedFlowSignal
=
300
. Set variable
KnownFlow
=
30.0
.
b. Start the calibration by setting variable
CalMode
=
1
.
c. When
CalMode
increments to
3
, for the second point, set variable
SimulatedFlowSignal
=
550
. Set variable
KnownFlow
=
10
.
d. Resume the deployment calibration by setting variable
CalMode
=
4
4. When variable
CalMode
increments to
6
, the deployment calibration is
complete. Calibrated multiplier is
-0.08;
calibrated offset is
53.9
.
5. To continue this example, suppose the simulated sensor multiplier and offset
drift. Simulate a seven-day service calibration to correct the drift as follows:
a. Set variable
SimulatedFlowSignal
=
285
. Set variable
KnownFlow
=
30.0
.
b. Start the seven-day service calibration by setting variable
CalMode
=
1
.
c. When
CalMode
increments to
3
, set variable
SimulatedFlowSignal
=
522
. Set variable
KnownFlow
=
10
.
d. Resume the calibration by setting variable
CalMode
=
4
6. When variable
CalMode
increments to
6
, the calibration is complete. The
corrected multiplier is
-0.08
; offset is
53.9
.
FieldCal() Two-Point Slope and Offset
'This program example demonstrates the use of FieldCal() in calculating and applying a
'multiplier and offset calibration. A multiplier and offset calibration compares signal
'magnitudes of a sensor to known standards. The calculated multiplier and offset scale the
'reported magnitude of the sensor to a value consistent with the linear relationship that
'intersects known points sequentially entered in to the FieldCal() KnownVar parameter.
'Subsequent measurements are scaled by the new multiplier and offset.
'This program demonstrates the multiplier and offset calibration with the following procedure:
' -- Simulate a signal from a flow sensor.
' -- Measure the 'sensor' signal.
' -- Calculate and apply a multiplier and offset.
'You can set up the simulation by loading this program into the CR800 and interconnecting
'the following terminals with a jumper wire to simulate a flow sensor signal as follows:
'
Vx1
---
SE
1
'For the simulation, the value of the calibration standard and the initial 'sensor' signal
'are set automatically. Start the multiplier-and-offset routine by setting variable
'CalMode = 1. The value in CalMode will increment automatically. When CalMode = 3, set
'variables SimulatedFlowSignal = 550 and KnownFlow = 10, then set CalMode = 4. CalMode
'will again increment automatically. When CalMode = 6 (occurs automatically after 10
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......