Section 7. Installation
259
This manual includes this discussion of PRTs because of the following:
•
Many applications need the accuracy of a PRT.
•
PRT procedures confuse many users.
•
PRTs are not usually manufactured ready to use for most CR800 PRT
setups.
This section gives procedures and diagrams for many circuit setups. It also has
relatively simplified examples of each circuit type and associated CRBasic
programming.
7.7.16.1 Measurement Theory (PRT)
RTDs (resistance temperature detectors) are resistive devices made of platinum,
nickel, copper, or other material. Platinum RTDs, known as PRTs (platinum
resistance thermometers) are very accurate temperature measurement sensors.
This discussion focuses
on the 100 Ω PRT. Apply the following principles to
other RTDs.
•
A PRT element is a specialized resistor with two connection points.
Most PRTs are either 100 Ω or 1000 Ω. This number is the resistance
the PRT has at 0 °C.
•
The resistance of a PRT increases as it is warmed. Industrial standards
define how PRTs respond to temperature; see
PRT Callendar-Van Dusen
Coefficients
•
There are many ways to measure a PRT with a CR800 datalogger.
When using
Vx
terminals , the most direct route is to measure a four-
wire PRT in a three-wire half bridge. Other ways to measure a PRT are
listed in
TABLE: PRT Measurement Circuit Overview
•
Better excitation accuracy results if the highest possible excitation is
used. Better measurement resolution results if the voltage output range
from the PRT spans the analog-input voltage range of the CR800.
Better measurement accuracy occurs when the output signal can be kept
as large as possible. Procedures in the following example balance these
best practices.
•
A feature of PRT measurements is the ratio RS/RS
0
, where RS is the
PRT resistance now and RS
0
is the PRT resistance at 0 °C. RS/RS
0
makes it easy to apply the results of an ice-bath calibration to a
temperature measurement. For jobs that do not need veryhigh accuracy,
skip the calibration and assume that PRT resistance at 0 °C is either 100
Ω or
1000 Ω.
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......