Section 8. Operation
343
Calibrate()
Instruction Results
Cal()
Array
Field
Descriptions of Array Elements
Typical Value
Differential (Diff)
Single-Ended (SE)
Offset or Gain
±mV Input
Range
Integration
52
SE
Offset
2.5
50 Hz Rejection
±20 LSB
53
Diff
Offset
2.5
50 Hz Rejection
±20 LSB
54
Gain
2.5
50 Hz Rejection
–0.00067 mV/LSB
8.1.2.5 Strain Measurements — Details
Related Topics:
•
Strain Measurements — Overview
(p. 70)
•
Strain Measurements — Details
(p. 343)
•
FieldCalStrain() Examples
(p. 228)
A principal use of the four-wire full bridge is the measurement of strain gages in
structural stress analysis.
StrainCalc()
calculates microstrain (
µ
ε) from the
formula for the particular strain bridge configuration used. All strain gages
supported by
StrainCalc()
use the full-bridge schematic. In strain-gage parlance,
'quarter-bridge', 'half-bridge' and 'full-bridge' refer to the number of active
elements in the full-bridge schematic. In other words, a quarter-bridge strain
gage has one active element, a half-bridge has two, and a full-bridge has four.
StrainCalc()
requires a bridge-configuration code. The table
StrainCalc()
Instruction Equations
(p. 343)
shows the equation used by each configuration code.
Each code can be preceded by a dash (
-
). Use a code without the dash when the
bridge is configured so the output decreases with increasing strain. Use a dashed
code when the bridge is configured so the output increases with increasing strain.
A dashed code sets the polarity of V
r
to negative.
StrainCalc()
Instruction Equations
StrainCalc()
BrConfig
Code
Configuration
1
Quarter-bridge strain gage
1
:
2
Half-bridge strain gage
1
. One gage parallel to strain, the
other at 90° to strain:
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......