Section 7. Installation
217
4. Set
KnownVar
variable to the offset or zero value.
5.
Set mode variable =
1
to start calibration.
7.7.11.4.2 Two-Point Calibrations (gain and offset)
Use this two-point calibration procedure to adjust multipliers (slopes) and offsets
(y intercepts). See
FieldCal() Slope and Offset (Opt 2) Example
(p. 223)
and
FieldCal() Slope (Opt 3) Example
(p. 225)
for demonstration programs:
1. Use a separate
FieldCal()
instruction and separate variables for each sensor to
be calibrated.
2.
Ensure mode variable =
0
or
6
before starting.
a. If
Mode
>
0
and ≠
6
, calibration is in progress.
b. If
Mode
<
0
, calibration encountered an error.
3. Place sensor into first known point condition.
4. Set
KnownVar
variable to first known point.
5. Set
Mode
variable =
1
to start first part of calibration.
a.
Mode
=
2
(automatic) during the first point calibration.
b.
Mode
=
3
(automatic) when the first point is completed.
6. Place sensor into second known point condition.
7. Set
KnownVar
variable to second known point.
8. Set
Mode
=
4
to start second part of calibration.
a.
Mode
=
5
(automatic) during second point calibration.
b.
Mode
=
6
(automatic) when calibration is complete.
7.7.11.4.3 Zero Basis Point Calibration
Zero-basis calibration (
FieldCal()
instruction
Option 4
) is designed for use with
static vibrating wire measurements. It loads values into zero-point variables
to track conditions at the time of the zero calibration. See
FieldCal() Zero
Basis (Opt 4) Example
(p. 228)
for a demonstration program.
7.7.11.5 Field Calibration Examples
FieldCal()
has the following calibration options:
•
Zero
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......