Section 5. Overview
67
FIGURE 19: Analog Sensor Wired to
Differential Channel #1
Differential and Single-Ended Input
Terminals
Differentiaol
DIFF Terminals
Single-Ended
SE Terminals
1H
1
1L
2
2H
3
2L
4
3H
5
3L
6
5.2.2.1.1 Single-Ended Measurements — Overview
Related Topics:
•
Single-Ended Measurements — Overview
(p. 67)
•
Single-Ended Measurements — Details
(p. 350)
A single-ended measurement measures the difference in voltage between the
terminal configured for single-ended input and the reference ground. While
differential measurements are usually preferred, a single-ended measurement is
often adequate in applications wherein some types of noise are not present and
care is taken to avoid problems caused by
ground currents
(p. 501).
Examples of
applications wherein a single-ended measurement may be preferred include:
•
Not enough differential terminals available. Differential measurements
use twice as many
H
/
L
terminals as do single-ended measurements.
•
Rapid sampling is required. Single-ended measurement time is about half
that of differential measurement time.
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......