4. Select File Control
at the top of the Connect window.
5. Click Send
at the top of the File Control window.
6. Navigate to the C:\Campbellsci\Lib\OperatingSystems folder.
7. Ensure Datalogger Operating System Files (*.obj) is selected in the files of type list, select
the new OS .obj file, and click Open to update the OS on the data logger.
Note the following precautions when sending as a program:
l
Any peripherals being powered through the SW12 terminals will be turned off until the
program logic turns them on again.
l
Operating systems are very large files. Be cautious of data charges. Sending over a direct
serial or USB connection is recommended, when possible.
A.3 CRVW3 maintenance with
LoggerNet
From time to time, Campbell Scientific will release updates to the CRVW3 firmware, and make
these available for download from the Campbell Scientific website. See
(p. 72) for details about how to update the CRVW3 firmware, including remote updates
made from within
LoggerNet
.
If the measurement output of the CRVW3 has been configured for engineering output and
baseline readings are used as part of that calculation, the sensors may need to be periodically
calibrated using the
Calibration Wizard
. See
Engineering output and calibration
(p. 81) for details
about calibration and using the
Calibration Wizard
.