CS120A and CS125 Visibility and Present Weather Sensors
36
12 0 0 1 0 92 M1 000000000000 2 0 3 0 92 135 88EF
This would give the following output:
Above the custom fields are “1 000000000000 2 0 30 92 135”. The averaging
duration is 1 minute, no system alarms are set, the emitter is reading 2%
contamination, the detector is reading 0% contamination, the SYNOP code is
30, the 10 minute average MOR is 92 m and the TMOR is 134 m. The
following shows the number of values output.
List of custom output values
Field
Description
Number of
values output
1
Averaging duration
1
2
User alarms
2
3
System alarms
12
4
Dirty windows values, as percentages, emitter followed by detector
2
5
Sensor serial number
1
6
Particle count
(1)
, particles in the previous minute
1
7
Intensity
(1)
, mm/hr
1
8
Accumulation
(1),(2)
1
9
Generic SYNOP
(1)
1
10
SYNOP code
(1)
1
11
METAR code
(1)
1
12
NWS code
(1)
1
13
Temperature (degrees C)
(1)
1
14
Humidity (%)
(1)(3)
1
15
Visibility averaged over the last 10 minutes
1
CS125 MESSAGE - SUB 4
Configure the message output format:
1 - Averaging duration = 1
2 - User alarms = 0
3 - System alarms = 1
4 - Dirty windows values = 1
5 - Sensor serial number = 0
6 - Particle count = 0
7 - Intensity = 0
8 - Accumulation = 0
9 - Generic SYNOP = 0
10 - SYNOP code = 1
11 - METAR code = 0
12 - NWS code = 0
13 - Temperature (Degrees C) = 0
14 - Humidity (%) = 0
15 – 10 minute ave. vis. = 1
16 – TMOR = 1
18 – Past SYNOP = 0
19 – EXCO = 0
Or, type 0 to exit without changes.
->