CSAT3B Three-Dimensional Sonic Anemometer
3
FIGURE 4-2. CSAT3B mounting
4.
Ground the CSAT3B to the tower by attaching a user-supplied, heavy-
gauge wire from the copper grounding lug on the back of the CSAT3B
block (FIGURE
).
5.
Earth (chassis) ground the other end of the wire to the CSAT3B mounting
structure or to a grounding rod. For more details on grounding, see the
grounding section of the CR3000 data logger manual.
If connecting multiple CSAT3Bs together either in a series or a
star topology, each CSAT3B must be separately grounded to
either the mounting structure or a grounding rod.
FIGURE 4-3. Grounding lug of CSAT3B
NOTE
CM20X Crossarm
CM250 Leveling Mount
CSAT3B Grounding Lug