Appendix A. Normalization and Function Test
A-3
1.
Click on “Auto
-
detect Sensors” in the lower left corner. Wait until all
sensors are auto-detected. If the number of sensors detected does not
correspond with the number of sensors on the probe, check to ensure that
each sensor is addressed sequentially, beginning with address 1 on the top
sensor.
2.
Enter the depth of each sensor in the “Depth” column. Each depth will be
a multiple of 10.
3.
Enter “1.000000;1.000000;0.000000”
on each line in the “Equation
A;B;C” column.
4. Hold the probe in the air, away from any other object, with your hands
away from the sensors, i.e. set the end of the probe on the floor and hold it
upright. For each sensor, press the “High / Air” toggle b
utton
. Allow
about 10 seconds, then toggle the button off. Values will be within a few
thousand of those shown in the example above.
5. Fill the normalization chamber (P/N SEN70056) or a deep bucket with
water. If using the normalization chamber, take the probe out of the
access tube. For each sensor, slide the probe into the normalization
chamber tube such that the sensor is in the center of the tube. If using a
bucket, submerge most of the white tube portion of the probe into the
center of the bucket. Do not submerge the blue portion of the probe, or
allow the electronics to get wet. Press the “Low / Water” toggle button
. Allow about 10 seconds, then toggle the button off. Values will be
within a few thousand of those shown in the example above.
6. Set the SDI-12 probe address in the lower right area of the IPConfig
window. The SDI-12 address is used in the datalogger program. If only
one probe is to be used on a datalogger SDI-12 input channel, you may
leave the address at “0.”
7.
Press “Write to Probe” in the lower right of center of the window.
8.
Press “Backup Configuration” in the lower left of center of the window.
Use the default file name.