Table 10-4: ConfigCmd Values for Setting and Retrieving Settings
ConfigCmd
variable
Setting description (some settings list possible values for the
DestSource
variable)
Set
Retrieve
9
109
Unprompted Output Mode: 0 = Disable, 1 =USB, 2 = RS-485
10
110
RS-485 Baud Rate
11
111
Span/Zero Control: 0 = Inactive, 1 = Zero, 2 = Span CO
2
, 3 = Span
H
2
O (see
ConfigCmd 11 zero-and-span control
12
112
CO
2
Span Concentration
13
113
H
2
O Span Dewpoint Temperature
14
114
CO
2
Zero Coefficient
15
115
CO
2
Span Coefficient
16
116
H
2
O Zero Coefficient
17
117
H
2
O Span Coefficient
18 or 218**
118
Heater Voltage (0 to 4.0V, −1 = Off) (see
20
120
Analog Output Enable: 0 = Disable, 1 = Enable
21
121
PowerDown: 0 = Gas Head On, 1 = Gas Head Off
22
122
Unprompted Output Rate: 10 = 10 Hz, 20 = 20 Hz, 99 = motor-spin
rate
24
124
EC155 Pressure Sensor Type: 0 = Differential, 1 = Absolute
25
125
Sonic Transducer Shadow Correction: 0 = Disable, 1 = Enable
99
N/A
Save Settings to EEPROM memory
10.3.1 ConfigCmd 11 zero-and-span control
To perform zeroing of CO
2
and H
2
O,
ConfigCmd 11
is set to 1. After the IRGASON completes
the zero, it will write the value to -1. The data logger program can poll this value or simply wait for
a period of time to allow the zeroing to complete. To perform CO
2
span, the CO
2
Span
Concentration
setting (
ConfigCmd 12
) must be written to the proper value in ppm CO
2
prior to setting the zero-and-span control setting (
ConfigCmd 11
) to 2. After the CO
2
span is
completed, the value of the zero-and-span control setting will change to -2. H
2
O span is similar
to CO
2
. First the
H
2
O Span Dew Point Temperature
value (
ConfigCmd 13
) must be
IRGASON® Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
51
Summary of Contents for IRGASON
Page 1: ...Revision 08 2021 Copyright 2010 2021 Campbell Scientific Inc Product Manual ...
Page 73: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 69 ...
Page 74: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 70 ...
Page 75: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 71 ...
Page 77: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 73 ...
Page 78: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 74 ...
Page 79: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 75 ...
Page 80: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 76 ...
Page 81: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 77 ...
Page 82: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 78 ...
Page 83: ...D 3 Decarbite IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 79 ...
Page 84: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 80 ...
Page 85: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 81 ...
Page 86: ...IRGASON Integrated CO2 H2O Open Path Gas Analyzer and 3D Sonic Anemometer 82 ...