LI200RX Pyranometer
13
Symptom: Incorrect solar radiation
1. Make sure the top surface of the sensor head is clean, and that the sensor
is properly leveled.
2. Verify that the range code, multiplier, and offset parameters are correct
for the desired engineering units and datalogger type.
8.2 Maintenance
On a monthly basis, the level of the pyranometer should be checked. Keep the
sensor clean and treat it as a scientific instrument to maintain the accuracy of
the calibration. The vertical edge of the diffuser must be kept clean to maintain
appropriate cosine correction.
Clean the sensor only with water and/or a mild detergent such as dishwashing
soap. Vinegar can also be used to remove hard water deposits from the diffuser
element if necessary.
Do not use alcohol, organic solvents, abrasives, or strong
detergents to clean the diffuser element. Exposure to
alcohol or organic solvents harm the acrylic materials of the
sensor, which will adversely affect the cosine response of
the sensor.
8.3 Calibration
Recalibrate the LI200RX every two years.
Send in the entire sensor (not just the head) for recalibration.
LI200RX pyranometers output a current that is proportional to the incoming
solar radiation. Each LI200RX has a unique calibration factor. A variable shunt
resistor in the cable converts the current to the voltage measured by the
datalogger. Campbell Scientific sets the shunt resistor so that the pyranometer
outputs 5 mV kW
–1
m
2
.
The resistor value is found using Ohms law. The resistance is found by
dividing the desired output voltage by the calibrated current output. For
example, a pyranometer with a calibration of 92
µ
A kW
–1
m
2
, will have the
resistor set to:
54.35
Ω =
−
−
5
0 092
1
2
1
2
mV kW m
mA kW m
.
9. Acknowledgements
Campbell Scientific, Inc. gratefully acknowledges the contribution of LI-COR
®
to concepts, text, and images used in this manual.
CAUTION
CAUTION
Summary of Contents for LI200RX
Page 2: ......
Page 6: ......
Page 22: ...LI200RX Pyranometer 14 ...
Page 24: ......
Page 27: ......