PVS5120-Series Portable Samplers
42
examples for the CR1000 can be found on our website at
and in Appendix
(p. B-4)
TABLE 8-5. VSC100 Public Table for PakBus® Control
Array
Value
Name
Description
1
Bottles
Number of bottles in sampler.
2
SampPerBottle
Number of samples to be put in each bottle(s).
3
NeverFull
Determines if the sampler loops without stopping
(must be manually serviced before the sampler is
full to prevent overflow).
4
PurgeTime1
Initial purge time, in seconds, that the sampler
hose is purged before collecting sample.
5
LevelSetTime Time, in seconds, to evacuate sample chamber to
desired sample volume.
6
DepositTime
Time in seconds to hold pinch valve open to
allow sample to be deposited in bottle.
7
PurgeTime2
Final purge time, in seconds, that the sampler
hose is evacuated after collecting sample.
8
UsePurgeTime
A true/false variable which indicates if
PurgeTime2 is set independently of PurgeTime1
or is set to same value as PurgeTime1.
9
SetTimeOut
True/false value that indicates whether the
vacuum time is set by the user or automatically
set (2 x PurgeTime1).
10
TimeOut
Vacuum timeout, in seconds, to determine how
long sampler should attempt to collect a sample
before triggering an error.
11
SamplerType
The sampler type determined by the
corresponding integer value:
0 = PakBus, 1 = Pulse, 2 = Modbus,
3 = SDI-12, 4 = Time
12
Bottle
Indicates bottle number that the sample is
deposited in (1 to 24). Value is always 1 for
composite sampler. (Read only.)
13
SampleCount
Displays the number of samples deposited in to
this bottle. Value is always 1 for never full
option. (Read only.)
14
Response
Displays numeric code ranging from 0 to 8 that
represents a different condition (TABLE
). 0
indicates no errors. (Read only.)
15
Batt
Returns the sampler’s battery voltage just before
that sample was collected. (Read only.)
Summary of Contents for PVS5120 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 70: ...Appendix A Sample Transport Velocity A 2...
Page 78: ...Appendix B Example Programs B 8...
Page 86: ...Appendix C Monitoring Sampler Status via RS 485 C 8...
Page 96: ...Appendix G Generic Modbus Control G 4...
Page 100: ...Appendix H ISO5667 Conformity H 4...
Page 101: ......