Appendix B. Calibration Wizard Tutorial
The Calibration Wizard is designed to help users:
•
Calibrate TDR probes to eliminate any probe to probe variability
•
Create probe-specific lengths and offsets for user-made probes
•
Quickly determine the appropriate cable and window lengths of any
probe
•
Calibrate the electrical conductivity parameters (probe cell constant)
Before using the Calibration Wizard, physically connect all
multiplexers and sensors the way they will be permanently
deployed. This will ensure that the Calibration Wizard determines
all measurement parameters correctly.
1. To start the Calibration Wizard, select
.
2. Select whether to
Calibrate Probe Length and Probe Offset
only or to
also
Calibrate Electrical Conductivity Parameters
. Click
Next
.
3. Place probe in deionized water with at least 3 cm clearance from the sides
and then click on
Next
to determine the appropriate cable length and
window length.
NOTE