Campbell Scientific suggests doubling the settling time every 50 ft. The following code snip
examples increase settling time by 20000 µs by placing
20000
as the argument in the
SettlingTime
parameter:
CR6-series data logger example:
Therm109
(T109_C,1,U1,U10,
20000
,60,1.0,0.0)
CR800-series and CR3000 data loggers example:
Therm109
(T109_C,1,1,1,
20000
,_60Hz,1.0,0.0)
9. Troubleshooting and
maintenance
NOTE:
All factory repairs and recalibrations require a returned material authorization (RMA) and
completion of the “Declaration of Hazardous Material and Decontamination” form. Refer to
the
page at the back of this manual for more information.
9.1 Troubleshooting
Symptom: Temperature is reported as NAN, –INF, or incorrect temperature.
Verify wires are connected to the terminals specified in the
Therm109()
instruction: red to
single-ended analog input (SE or U), black to switched excitation (VX/EX or U), and purple to
ground (
⏚
).
Symptom: Incorrect temperature is reported.
Verify the
Mult
and
Offset
arguments in
Therm109()
are correct for the desired units
(
(p. 6)). Check the cable for signs of damage and possible moisture
intrusion.
109SS Temperature Probe