background image

Cut

 is an easy way to program your data logger to measure the sensor and assign data logger 

wiring terminals. 

Short Cut

 is available as a download on 

www.campbellsci.com

 

. It is included 

in installations of 

LoggerNet

RTDAQ

, and 

PC400

.

The following procedure also describes programming with 

Short Cut

.

 1.  Open 

Short Cut

 and click Create New Program.

 2.  Double-click the data logger model.

 3.  In the Available Sensors and Devices box, type 109 or find the 109 in the Sensors > 

Temperature folder. Double-click the 109 Temperature Probe. Data defaults to degree 
Celsius. This can be changed by clicking the Deg C box and selecting Deg F, for degrees 
Fahrenheit, or K for Kelvin. 

109SS Temperature Probe

     

2

Summary of Contents for TEMPERATURE PROBE 109SS

Page 1: ...Revision 05 2021 Copyright 1983 2021 Campbell Scientific Inc...

Page 2: ...oil temperature installation 8 8 Operation 8 8 1 Sensor schematic 8 8 2 Measurement and output linearization 8 8 3 Electrically noisy environments 9 8 4 Long cable lengths 9 9 Troubleshooting and main...

Page 3: ...ning when tested according to U L 94 H B and passes FMVSS302 However local fire codes may preclude its use inside buildings 3 Initial inspection Check the packaging and contents of the shipment If dam...

Page 4: ...ng procedure also describes programming with Short Cut 1 Open Short Cut and click Create New Program 2 Double click the data logger model 3 In the Available Sensors and Devices box type 109 or find th...

Page 5: ...wired to the data logger Click OK after wiring the sensor 5 Repeat steps three and four for other sensors being measured Click Next 6 In Output Setup type the scan rate Data Output Storage Intervals...

Page 6: ...s a rugged probe that accurately measures soil or water temperature in a variety of applications The sensor consists of a thermistor encased in a stainless steel sheath This design protects the thermi...

Page 7: ...ble length 305 m 1000 ft Accuracy1 Worst case 0 60 C 40 to 70 C 0 49 C 20 to 70 C Maximum submergence 45 m 150 ft 63 psi Interchangeability error 0 60 C at 40 C 0 38 C at 0 C 0 10 C at 25 C 0 30 C at...

Page 8: ...Data logger programming p 6 Short Cut does this work for you See QuickStart p 1 for a Short Cut tutorial 7 1 Wiring to data logger Table 7 1 Wire color function and data logger connection Wire color...

Page 9: ...vided in Measurement and output linearization p 8 7 2 1 Therm109 instruction The Therm109 measurement instruction programs CRBasic data loggers to measure this sensor It applies a precise excitation v...

Page 10: ...ffect of solar heating of the above ground cable on the temperature measurement Placement of the cable inside a rugged conduit may be necessary for long cable runs especially in locations subject to d...

Page 11: ...ficients provided by Measurement Specialties 8 3 Electrically noisy environments EMF noise emanating from the ac mains power grid can be a significant source of measurement error 60 Hz noise is common...

Page 12: ...A and completion of the Declaration of Hazardous Material and Decontamination form Refer to the Assistance page at the back of this manual for more information 9 1 Troubleshooting Symptom Temperature...

Page 13: ...n If needed an estimate of the interchangeability error for 0 to 50 C that can be used as the Offset parameter of the Therm109 instruction can be determined with a 1 point or 2 point calibration Calib...

Page 14: ...ted for additional refinement NOTE Once the file is edited with CRBasic Editor Short Cut can no longer be used to edit the program it created 2 To add the Short Cut wiring information into the new CRB...

Page 15: ...Wire Data logger Color Function Terminal Black Voltage excitation input VX1 EX1 Red Analog voltage output SE1 Purple Bridge resistor ground Ground Symbol Clear Shield Ground Symbol Declare the variabl...

Page 16: ...40 336103 2 40 00 39 314558 39 00 38 294529 1 38 00 37 275900 8 37 00 36 258567 36 00 35 242430 2 35 00 34 227400 9 34 00 33 213396 6 33 00 32 200341 4 32 00 31 188165 5 31 00 30 176804 8 30 00 29 166...

Page 17: ...17 81641 4 17 00 16 77138 6 16 00 15 72911 1 15 00 14 68940 4 14 00 13 65209 7 13 00 12 61702 9 12 00 11 58405 5 11 00 10 55303 9 10 00 9 52385 2 9 00 8 49637 8 8 00 7 47050 6 7 00 6 44613 4 6 00 5 4...

Page 18: ...00 10 19903 2 10 00 11 18973 3 11 00 12 18092 2 12 00 13 17256 9 13 00 14 16464 9 14 00 15 15713 7 15 00 16 15000 9 16 00 17 14324 5 17 00 18 13682 3 18 00 19 13072 6 19 00 20 12493 3 20 00 21 11943...

Page 19: ...5 6530 3 35 00 36 6266 2 36 00 37 6014 3 37 00 38 5773 8 38 00 39 5544 2 39 00 40 5325 40 00 41 5115 6 41 00 42 4915 6 42 00 43 4724 4 43 00 44 4541 7 44 00 45 4367 45 00 46 4200 46 00 47 4040 2 47 00...

Page 20: ...3 9 58 00 59 2578 6 59 00 60 2487 1 60 00 61 2399 4 61 00 62 2315 2 62 00 63 2234 4 63 00 64 2156 8 64 00 65 2082 3 65 00 66 2010 8 66 00 67 1942 1 67 00 68 1876 68 00 69 1812 6 69 00 70 1751 6 70 00...

Page 21: ...ther manufacturer s products that are resold by Campbell Scientific are warranted only to the limits extended by the original manufacturer Refer to www campbellsci com terms warranty for more informat...

Page 22: ...ion or Repair Reference number contact your CAMPBELL SCIENTIFIC regional office Please write the issued number clearly on the outside of the shipping container and ship as directed For all returns the...

Page 23: ...rohibit climbing by other persons Take reasonable precautions to secure tripod and tower sites from trespassers l Use only manufacturer recommended parts materials and tools Utility and Electrical l Y...

Page 24: ...nfo campbellsci fr www campbellsci fr Germany Location Phone Email Website Bremen Germany 49 0 421 460974 0 info campbellsci de www campbellsci de India Location Phone Email Website New Delhi DL India...

Reviews: